Sensor Calibration
Calibration Using Calibration Panel CP 40
Page 35 (44)
at
os
2-
3_r
ev02_
so
-40
0-
800
-v7
-5
-0_en_r
ev
-b
2012
-01
-03
Calibration object at measuring distance to sensor
9.3.4
Calibration Process
After that the software will guide you through the individual calibration
steps. In the first step, you are asked to place the sensor and the cali-
bration object in the center of the measuring volume.
Enable the
Automatic exposure time
.
In exceptional cases it might happen that the function
Automatic ex-
posure time
cannot work correctly.
In this case, disable the function and use the exposure time wheel to
adjust the optimum exposure time for the camera image.
Exposure time adjustment for calibration