B l u e C o n n e c t – C o n f i g u r i n g t h e M o d u l e s
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 13 / 29
www.go-sys.de
4.4 The Info Window
After the program has detected the connected module (here Redox/ORP), the module info window opens.
4.5 The Calibration Window
A calibration compares the
value pairs
of the measured
sensor raw values
∗
and allocated
reference values
from calibration liquids. These value pairs are taken as points in a coordinate system. The curve of a 1. to 5.
Order polynomial is placed through these points as accurately as possible, this is how the calibration polyno-
mial is created.
Example with a 2. Order polynomial:
∗
A raw sensor value is the uncalibrated sensor measurement value or the uncalibrated current input value.
COM 1 selected
Serial Communication
Modbus
BlueConnect Redox
Info
Parameter
Calibration
Measuring
Data Processing
Data
Device
Firmware Version
Serial Number
Modbus Slave ID
Baudrate
Production Date
BlueConnect Redox
2.12
99
1
9600
25.10.2021
File Language
Modbus Tool V1.07
Calibration table
Calibration coefficients