Intensity
Display and setting of the number of light flashes per single measurement
as in
9.2.1 The Configuration Window of the Spectrometer
9.2.2.5 Base Bar Functions
Cleaning is running.
Measurement or intensity calibration in clearwater is running.
Transmission of spectra data to the spectrometer unit is running.
Clearwater calibration is running.
Please execute clearwater calibration.
Clearwater calibration is necessary, i.e. the calibration interval
1
of
the clear water calibration has been exceeded.
T=
nn
,
n
°C
Temperature of the spectrometer electronic
D=
nnn
MVR
2
of the dark measuring, see
7 Measurement Cycle
9.2.2.6 Fingerprint (previous function)
up to AMS software version 4.3.4.0 and BlueBox firmware version 4.02.14 only
The fingerprint is an additionally stored absorbance spectrum. It is saved via the drop-down menu under
<Application>, clicking on <Save fingerprint> saves the absorbance spectrum of the last measurement under
the designation fingerprint. The values of this fingerprint can be queried with AMS formula
3
, e.g. as reference
spectral values.
1
see
9.2.1 The Configuration Window of the Spectrometer
there
Max. calibration interval [days]
2
MVR
= Maximum digital Value of a Raw spectrum
3
see
Manual BlueBox PC Software
and
13.1 Example Calculation of a Fingerprint Difference
Saves the current spectrum as fingerprint.