CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION
Cleaning filter
USE PROTECTIVE GLOVES WHEN MANIPULATING THE FILTERS AND THE DRAINS.
WASTE DISPOSE AND CONTAMINATED PRODUCTS IN GARBAGE BIOLOGICAL.
The vacuum pump filter should be cleaned in each 15 days. Before cleaning, you should
make the sample suction (see solution preparation in the previous item).
To clean the filter, follow the procedure below:
- After the solution suction
through the Suctorian, remove
the Biovac Sec covers that
access to the conduit head (1).
- Release the cable clip (2).
2
1
DAILY PROCEDURE FOR INTERNAL CLEANING OF SUCTION DUCTS
OF THE DRY VACUUM PUMP
(Hoses and internal parts of the Vacuum Pump ducts)
• After each patient leaves, perform the suction of 250 ml of clean water in each totally
open suctor;
• At the end of the working Day, perform the suction of the 250 ml mixture of bleach
mixed with 250 ml of clean water (proportionally divided in each used suctor);
WARNING:
• Never use foamy products in the suction (deep cleaner, detergents, floaters, etc), this
procedure may damage the internal parts of the dry vacuum pump’s engine;
• Never use the bleach solution for external cleaning of any equipment, because this
mixture is highly corrosive and may damage metal parts.