Doc. No. 0127420 rev. B
Page 14 of 15
Place the Xrec tube into the test fixture. Mount the snap
on adaptor on the coupler and place the XRec into it.
Connect the hearing instrument to the other end of the
Xrec tube
Test fixture p/n 16229100
Test equipment
Use XRec test tube p/n 16185100 when
testing
The test fixture must be adjusted before use. Loose the
two screws. The smallest part must be positioned so that
the test tube is in the center of the green circle. Fasten
the screws.
This adjustmest must also be done on a frequently basis
to ensure optimal testing.
Insert battery pill and flex strip into the HI
Snap-on adaptor p/n 15238500. To be used with Otometrics 2cc coupler.