GMI PS200 series Quick Operating Instructions Download Page 2

WARNINGS

 LOW BATTERY

Flashes when approximately 30 minutes operating time remains. The display
flashes red, the audible alarm sounds once every two seconds and the red
LED’s flash.  Note: Gas alarms will continue to operate as normal.

 BAT  FAULT

Flashes when approximately 3 minutes operating time remains. The display
will be red, the audible alarm sounds continuously and the red

 

LED’s remain

on. After approximately 3 minutes the instrument automatically switches off.

CAL EXPIRED

Displayed every 30 seconds if the calibration expiry date has been exceeded.

ZERO FAULT

 (and flashing 

 symbol)

Appears after warm-up if the instrument is switched on in gas or the instrument
has been unable to zero all sensors. The display will be red, the audible alarm
sounds once every 2 seconds, and the red

 

LED’s flash (start-up only). However,

the instrument can still be used to detect and alarm on the other sensor(s)
fitted.

Note:

1)   If a flashing 

 symbol is alternating with a gas value, leave instrument on

for 30 minutes then switch OFF, and then ON again in fresh air. If the fault
remains, return instrument to a GMI approved Service / Repair facility.

2)  If the 

 symbol is displayed constantly, then the sensor may require

replacement. Return instrument to a GMI approved Service / Repair facility.

 

 FLOW FAULT

 (Pumped instruments only.)

Flashes to indicate a sample fault . The display will be red, the audible alarm
sounds continuously and the red

 

LED’s remain on. Check sample line / filters

/ and probe for blockage. Clear the blockage and restart pump by a press and
hold of the 

 button.

-----------------------------------------------------------------   Part Number: 64191   Issue 3    07/07/2011   -----------------------------------------------------------------

SINGLE PRESS

PRESS & HOLD

Backlight ON

Manual Datalog

SWITCH OFF

&

SWITCH ON

Pump ON / OFF

Acknowledge / Mute Alarm

Reset MAX / MIN            

Switch ON - Manual Bump

Self Test

MAX

MIN

(

)

displayed

when

LOCATING SLOT

LOCATING TONGUE

CHARGING CLIP

USB CHARGING

CABLE

Refer to User Handbook for
alternative charging options.

CHARGING

Before first use the PS200 instrument should be charged for up to 4 hours or
until the ‘CHARGE COMPLETE’ message appears on the display.

The PS200 Series uses a Li-Ion (Lithium Ion) rechargeable battery. To
recharge, securely connect the charging clip to the bottom of the
instrument.Connect one end of the cable to the charging clip and the other to
the universal charging adaptor (supplied) or any standard USB socket.

BUTTON OPERATION