GMI PS1100 Quick Operating Instructions Download Page 2

To        CALb if calibration is required

CALb Background

Re-start Instrument

OFF

Switch OFF 

Fit for use

7

6

Press the button until the 

last LED is turned OFF

8

8

A

B

Fresh air 400ppm

C

Gas Measurement Instruments Ltd

www.gmiuk.com

 GMI Document Part Number: 63331  Issue 1 

When the instrument is in CALb 

mode, the first Background 

Calibration  is  done  after  five  (5) 

hours. If the instrument is left in 

CALb mode, it continues to do a 

Background Calibration every four 

(4)  hours  until  the  instrument  is 

restarted or switched OFF.

NOTE

If the status indicator is shown, the readings are not reliable!

The actions to take are as follows:

1) Charge the battery

2) Re-calibrate the carbon dioxide sensor zero point 

 

    (put the unit in CALb mode)

3) If the status indicator is still shown – contact your distributor

For logging of data and more detailed information, refer to the User 

Manual (GMI Part No. 63330).