45
Rev.05.15
Type
The urology programs require the use of specific endovaginal and endorectal
electrode probes, certified according to the Directive for Medical Devices
93/42/EEC. These are bipolar probes with a 2-mm female adapter which
attaches to 2-mm male cables.
Warnings
Urological electrostimulation is a medical application, which must be performed
under medical supervision.
Use
To use the probe electrode correctly, follow the instructions provided by the
manufacturer and given by the physician.
Maintenance
For cleaning, sterilization and disinfection, refer to the manufacturer's
instructions.
Suggestions
In case of deterioration of the probe electrode, replace it immediately.
Action Now program list
Action Now programs are normal EMS programs, with the only difference that each
single action will start only after pressing * button. The Action Now programs are
particularly useful to combine and synchronize the electric stimulation with a
voluntary action.
This program is particularly suggested in the sports field, for athletic preparation,
since the muscle contraction induced by the stimulator can be combined with a
dynamic or isometric workout.
Operating mode:
contraction will start after pressing * button. To interrupt
contraction before the end of contraction time, press * button again. In this case
the program will reset the rest period and will start at the beginning of the following
ramp, which can be activated by the user by pressing * again.
Action Now program list includes 7 parameter combinations (only for Genesy 600).
Area
Name
Hz
Ramp-Up time
Contraction time
Upper limbs
Action 0,2 - 1 s
30
0,2
1
Action 0,5- 1s
0,5
1
Lower limbs
Action 1 - 1 s
50
1
1
Action 2 - 1 s
2
1
Trunk
Action 3 - 2 s
80
3
2
Action 4 - 2 s
4
2
Action 2 - 6 s
100
2
6
TOTAL
84 programs
CE0476 does not refer to non-medical treatments.