The centrifuge rotor can separate samples whose density is lower than 2.0g/ml. If the separated sample’s density is over 2.0g/ml, please
calculate the allowable speed using the following formula:
Allowable Speed (rpm) = Maximum Speed × (2�0(g/ml) / Sample Density (g/ml))
1/2
GCC-MP / Multi-Purpose
Clinical Centrifuge
Quick Start Guide
Rotor
Tube
Adapter
Maximum
Maximum Centrifugal
Maximum
Type
Type
Required
Speed (rpm)
Radius (cm)
RCF (×g)
2.0/1.5mL x 30
—
6000
10
4020
0.2mL x 30
GCM-24-AD2
6000
8.5
3415
0.5mL x 30
GCM-24-AD5
6000
9
3618
2.0/1.5mL x 60
—
6000
10
4020
0.2mL x 60
GCM-24-AD2
6000
8.5
3415
0.5mL x 60
GCM-24-AD5
6000
9
3618
8-PCR x 12
—
6000
10
4020
0.2mL x 96
—
6000
10
4020
50mL x 6
—
6000
10.7
4306
15mL x 30
—
4500
R1 = 14.2
3210
R2 = 12.2
2760
15mL x 16
—
5000
12.4
3460
50mL x 8
—
5000
12.4
3460
100mL x 4
GCC-MP-SWING-100ML4
4000
15.9
2840
50mL x 4
GCC-MP-SWING-50ML8
4000
15.9
2840
10mL x 16
12mL x 16
GCC-MP-SWING-15ML16
4000
15.1
2700
15mL x 16
3mL x 36
7mL x 36
GCC-MP-SWING-7ML36
4000
14.7
2630
MTP
—
4000 12.1 2160
(128 x 85.6 x 15)
Cell Culture
—
4000 12.1 2160
(128 x 85.6 x 21)
DWP
—
4000 12.1 2160
(128 x 85.6 x 45)
Kits
—
4000 12.1 2160
(128 x 85.6 x 60)
GCC-MP-2ML60
GCC-MP-2ML30
GCC-MP-PCR
GCC-MP-50ML6
GCC-MP-15ML
GCC-MP-50ML8
GCC-MP-SWING
GCC-MP-SWING-
PLATE