Global Water WQ770 Manual Download Page 4

 

Global Water 

800-876-1172 

 globalw.com 

 

 

- 4 - 

 

 

III.

 

WQ770 Operation 

 

a.

 

Before turning the unit on, make sure that the sensor cable has 

correctly been attached to the display.  Push the cable connector 

gently into the meter’s connector.  The outer connector ring can be 

turned and locked into place so the cable will not detach from the 

meter during use. 

b.

 

Remove the battery cover from the meter and verify that a 9V lithium 

battery is installed in the unit.  Replace the battery cover. 

c.

 

Press the Menu/On Button. This brings up the display that shows the 

output from the turbidity sensor.   

d.

 

Press and hold the Menu/On button for 3 seconds to display the 

Range menu.  Use the Up/Down arrow buttons to select either the     

0-50 or the 0-1000 NTU range then press Save.   

e.

 

Press the Menu/On button again and use the Up/Down arrow keys to 

select units of NTU or ppm then press Save.  Pressing the Menu/On 

button 6 more times returns to normal display mode.  The other menu 

items are used for calibration and are described in Appendix A. 

f.

 

Place the sensor in the water to be tested.  The meter will display the 

turbidity of the water.  Warning:  Air bubbles in the water or under the 

sensor shield will cause errors in the readings.  These air bubbles 

must be avoided when placing the sensor in the water. 

g.

 

The Turbidity Meter will automatically shut off after 6 minutes.  To 

continue reading the display past this time press the Menu/On button 

prior to the meter shutting off. 

h.

 

Prolonged immersion in hot water will cause accelerated aging of the 

sensor optics and seals.  Warning:  Do not expose the sensor to 

water exceeding 50°C. 

 

 

 

 

IV.

 

Battery Life And Replacement 

 

Summary of Contents for WQ770

Page 1: ...alw com 1 Global Water Instrumentation Inc 151 Graham Road P O Box 9010 College Station TX 77842 9010 T 800 876 1172 Int l 979 690 5560 Fax 979 690 0440 E mail globalw globalw com Turbidity Meter WQ77...

Page 2: ...hat you will find the WQ770 to be a valuable asset for your application Should you require assistance our technical staff will be happy to help Table of Contents I Checklist Page 3 II Inspection 3 III...

Page 3: ...ipping there is no need to recalibrate the meter until the recommended maintenance period has elapsed II Inspection The Turbidity Meter was carefully inspected and certified by our Quality Assurance T...

Page 4: ...e 0 1000 NTU range then press Save e Press the Menu On button again and use the Up Down arrow keys to select units of NTU or ppm then press Save Pressing the Menu On button 6 more times returns to nor...

Page 5: ...he back of the meter The battery can be removed using your fingers or gently using a flat head screwdriver being careful not to damage the battery Ensure that the replacement battery is correctly inst...

Page 6: ...del 2 Unit serial number 3 P O the equipment was purchased on 4 Our sales order number or the invoice number 5 Repair instructions and or specific problems relating to the product Be prepared to descr...

Page 7: ...Global Water s obligations under this warranty are limited to at Global Water s option I replacing or II repairing any products determined to be defective In no case shall Global Water s liability exc...

Page 8: ...detector is placed at an angle of 90 degrees to the incident light beam measuring side scattered light The detected light intensity is directly proportional to the turbidity of the water b The followi...

Page 9: ...more beakers with 0 0 NTU Standard 50 NTU Standard and 1000 NTU Standard These beakers must be at least 500 ml in size or the end of the sensor will be too close to the bottom for accurate calibration...

Page 10: ...r bubbles from underneath the detector window ensuring that there are no bubbles on the lenses is critical Place the sensor into the distilled water Press the Save button until the display reads saved...

Page 11: ...in step b and set the HiLab value to 1000 in step i IX Appendix B Turbidity Standard Solutions The following turbidity standards are available from Global Water contact the Sales Department or Techni...

Reviews: