Chapter 1
Getting Started
S12V Service Manual
Page 1-1
S
COPE
This manual contains information with regards to application, installation, and operation of the
Silencer
®
S12V table-top centrifuge, and must be read prior to setup or operation of the
centrifuge.
U
NPACKING
A
ND
I
NSPECTION
Your new Silencer
®
centrifuge is packed in a specially designed shipping carton. Upon arrival
please check the condition of carton and note any damage. Keep all packing materials for the
warranty period of the centrifuge.
After unpacking your Silencer
®
carefully inspect it for damage. It is your responsibility to
report all damages to the transportation company and filing, if necessary, any damage claims.
At this point take a moment to ensure you have received the correct rotor, tube adapters, etc. If
not or if you have any questions regarding compatibility with your tubes, please contact GFMD,
Ltd. at 800-323-4306 before proceeding.
WARNING
Do not operate this equipment in a combustible
atmosphere!
S
ETUP
For your own safety and to prevent damage to the centrifuge, be sure to read this manual before
continuing with setup or operation of this centrifuge.
For smooth operation and long equipment life, choose a stable, level bench top.
Avoid sources of heat and/or moisture such as sinks, water baths, heating ducts, and direct
sunlight. A suitable environment is 10° to 30° C (50° to 86° F) with a relative humidity range of
40% to 70%. The minimum clearance is 6” on all sides. A cooler ventilated location and
leaving the lid open between cycles will allow longer and/or more frequent cycles without
overheating of samples. The maximum number of cycles that may be run consecutively will
depend upon operating speed, set time, ventilation around the centrifuge, and ambient
temperature. Samples that may be adversely affected by being warmed to 39°C during
centrifugation should be centrifuged in a refrigerated centrifuge such as the Silencer
®
S2510R.
Plug the power cord into a properly grounded outlet and turn on main power interrupt (on back
of unit next to power cord). For safety, equipment life, and proper performance, your Silencer
®
S12V’s power requirements are 115V AC, 5 amperes, 60Hz. The power cord must be UL listed
or CSA certified, 18 AWG, and 6’ 7” or less in length.