7
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Perform a Control Test:
Insert an Unused Strip
→
Make Sure the Code Number is Correct
→
Press “Front”
Button (3 secs)
→
Enter Control Solution Mode (Shown CL on Screen)
→
Shake the
Control Solution Well
→
Discard First Three Drops
→
Put One Drop onto a Clean
Surface
→
Touch the Control Solution with Strip Sample Inlet
→
Test Result Display
after Count Down
→
Compare the Result with the Range Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
Chapter 5
Control Solution Test
Control Range LV1
5.8-8.6 mg/dL
Storage &
Transportation
Storage &
Transportation
4
rip
BK-U1
7021
5.8-8.6 mg/dL
G1303190
2017/01
2017-01
EB13120901
2017-01
EB13120901