![Glas-Col LimiTrol 104A PC512 User Instructions Download Page 3](http://html1.mh-extra.com/html/glas-col/limitrol-104a-pc512/limitrol-104a-pc512_user-instructions_2230629003.webp)
J:\Std User Instructions\104A PC512, 104A PC524_LIMITROL II.doc 11/13
3
Specifications
Control Mode
-
Single input, single output
-
Range -210 to 1038 degrees C
Operator Interface
-
Control knob
-
Digital setpoint display
-
ON/OFF switch
-
Type J thermocouple receptacle
-
3-wire load receptacle
Input
-Type J thermocouple input; grounded or ungrounded
-
Automatic cold junction compensation and break protection for sensor.
Output
-7.5 amp, 120 Volt (Fused); 7.5 amp, 240 Volt (Push-to-reset circuit breaker)
Accuracy
- Calibration accuracy:
+ or - 1% of span.
- Set point accuracy:
+ or – 3% of span maximum.
Power
- 50/60 Hz
10%
-
Power consumption less than 4VA plus load.
Operating Environment
-
32 to 130 degrees F/ 0 to 55 degrees C; 0 to 90% RH, non-condensing