Glas-Col 104A PLSM112 Operating And Safety Instructions Manual Download Page 8

 

  

 

 

 

 

      Screen 1 

 

 

 

                 Screen 2 

 
To start auto-tuning: 

 

 

1.  Set the set point value with the arrow keys to the desired temperature of operation. 
2.  Press the 

Advance

 key until the Aut prompt appears in the right display. 

3.  Change the parameter to 

YES 

(displayed in left portion of display) with the arrow keys. 

4.  Press the 

Infinity

 key to display the set point and sensor temperatures. While the control is in the auto-

tuning mode, the display will alternate between the set point and the "Aut" prompt (Screen 1 & Screen 2). 
When tuning is complete, the lower display indicates only the set point. 

5.  When tuning is complete, the right display indicates setpoint only and Aut will revert to OFF. The control 

will use the auto-tuned PID values and retains them in the non-volatile memory. These values will be 
used on any further heating processes.  

6.  Repeat auto-tune process if heater/process is being changed. 

 

 

*To abort auto-tuning, the operator must reset Aut to OFF. Auto-tuning may also be aborted by cycling power off 
and on.  In all cases, aborting auto-tuning restores all original factory default values. 

 
Temperature overshoot 

 
In most cases this is due to a small sample size being heated with a high wattage heater. The control is set at the 
factory with the Power Scale High Output 1 (o.hi1) setting with a value of 100. This value can be adjusted in the 
setup menu, reference page 5 for instructions on how to access and navigate through the setup menu. If 
excessive temperature overshoot is occurring, adjust the value to 50 and start the heating process from ambient 
room temperature. If more than desired temperature overshoot is still occurring, decrease to a lower value and 
repeat heating process from ambient room temperature. If the proce

ss temperature doesn’t achieve the set point 

temperature or takes more than desired amount of time to reach set point temperature, increase the power scale 
high output 1 value. This process can take several value changes to obtain proper temperature control with your 
particular application setup. It is recommended to initiate the Auto-Tune process again after the proper power 
scale high output 1 value has been determined for your application.    

 

Error Code Definitions and Actions: 

 

Er.In - An open or reversed polarity sensor is the most likely cause. Check the sensor; if the connection is good 
and functions properly, call the factory.  Make sure the input parameter [SEn] is set to the correct type 
thermocouple. 
Er.Ab 

– Ambient temperature may be too hot or too cold. Make sure that the temperature surrounding the control 

is 

–18 to 65C. 

Er.CS 

– Checksum Error. Settings may have changed unexpectedly. Press the 

Infinity

 key to clear the error. 

Verify settings. If error message persists, contact the factory. 
 
To clear a corrected error, the power to the control needs to be cycled. 

GLOSSARY

 

 
Automatic prompts:  Data entry points where a microprocessor-based control "prompts" or asks the operator for 
information input. 
 

Summary of Contents for 104A PLSM112

Page 1: ...DIGITAL TEMPERATURE CONTROLS PLSM SERIES Models 104A PLSM112 104A PLSM112K 104A PLSM112T15A 120V 50 60 Hz 104A PLSM212 104A PLSM212K 104A PLSM212T15A 120V 50 60 Hz 104A PLSM312 104A PLSM312K 104A PLS...

Page 2: ...tch to the ON position Wait five seconds for the control unit to energize The control is shipped from the factory with the display reading in degrees C If a display in degrees F is desirable press the...

Page 3: ...Reset 0 00 to 99 99 repeats per minute Integral 0 00 to 99 99 minutes per repeat Rate or Derivative 0 00 to 9 99 minutes Operator Interface Advance Infinity Up and Down keys and ON OFF switch Dual fo...

Page 4: ...d Default 25 0 F or 14 0 C C Pb Cool Proportional Band Default 25 0 F or 14 0 C ti Time Integral Default 180 td Time Derivative Default 0 o tb1 Time Base Output 1 Default 1 0 or 20 0 o tb2 Time Base O...

Page 5: ...FF h Ag Heat Algorithm Default PID hSC Hysteresis Heat Cool Default 3 0 F or 2 0 C C Ag Cool Algorithm Default oFF A ty Alarm Type Default oFF A hy Alarm Hysteresis Default 1 0 A LA Alarm Latching Def...

Page 6: ...oftware Build Read Only Sn Serial Number Read Only dAtE Date of Manufacture Read Only USr r User Restore Settings nonE 2onE Zone Default oFF Chan Channel Default oFF CAL FCty Calibration Menu Operatio...

Page 7: ...e procedure isn t required in most applications Auto tuning This control can automatically tune the PID parameters to fit the characteristics of your particular thermal system Auto tune can only be us...

Page 8: ...instructions on how to access and navigate through the setup menu If excessive temperature overshoot is occurring adjust the value to 50 and start the heating process from ambient room temperature If...

Page 9: ...a set point by turning the output full On below set point and full Off above set point Output Action in response to difference between set point and process variable Overshoot Condition where temperat...

Page 10: ...apparatus due to improper installation or through attempts to operate the apparatus beyond its rated capacity intentional or otherwise Limitations of Warranties APART FROM SUCH WRITTEN STATEMENT OF W...

Reviews: