■ For your safety 

 

 

Make  certain  that  this  bathing  system  is  operated  only  by 
trained personnel. 

 

 

Always  check  the  water  temperature  before  the  client  enters 
the bathtub or when the shower spray is applied to the client. 

 

 

Make  certain  that  the  equipment  is  thoroughly  cleaned  after 
every use. 

 

 

Make sure that each client is assessed for it’s ambulatory ca-
pability, ensuring the client’s safety when entering and exit-
ing  the  bathing  system  (utilize  handgrips,  correct  entry 
height, correct client transferring device, etc.). 

 

 

When  using  disinfection,  wear  rubber  gloves  and  eye  protec-
tion. In case of eye or skin contact, follow the procedures on 
the disinfectant container and material safety data sheet. 

 

 

Always make sure that the device is supplied with power. 

 

 

Make sure that there are no materials right next to the tub to 
ensure a safe up and down movement. Disregarding this tip can 
cause bad injuries or damage the device. 

 

 

Don’t  deposit  any  things  on  the  bathtub.  Because  of  movement 
and vibrations, things can fall on the floor or in the tub. 

 

 

We  want  to  point  out  that  a  lifting  bathtub  is  a  nursing  tub 
and the bathing of the inhabitant has to be effected by use of 
a suitable lifter. 

 

 

Should  no  lift  be  used  we  recommend  to  furnish  the  bathtub 
with handles. 

 

 
 
Note:

  

The safety rules and instructions in this operating man-
ual must be observed in order to ensure safe and proper 
use  of  the  equipment.  Adjustments  and  repairs  must  be 
carried  out  by  authorized  and  appropriately  qualified 
persons. All claims under guarantee become void if these 
requirements are not met. 

For your safety 

Summary of Contents for GK Basic+

Page 1: ......

Page 2: ...iderably ease the strain involved in patient care Due to their advanced ergonomic design GK Bathtubs are simple and straightforward to use In spite of that we would ask you to study the instructions c...

Page 3: ...temperature 8 Shower 8 Filling the tub 9 Drain the tub 9 Disinfection 9 Flowmeter 9 Hydro massage 10 Fill in On 10 Fill in Off 10 Care cleaning of the tub 11 Maintenance Service 12 Intended use 13 Cer...

Page 4: ...sheet Always make sure that the device is supplied with power Make sure that there are no materials right next to the tub to ensure a safe up and down movement Disregarding this tip can cause bad inju...

Page 5: ...desired level help the client into the bathtub or bring the stretcher in the right position Use the UP or DOWN buttons to bring the tub in a comfort able working height If you want to use the shower a...

Page 6: ...Components 5 1 2 3 4 5 8 7 6 10 9 11 12...

Page 7: ...nents 6 Components 1 Disinfection 2 Shower 3 Stop valve fill in 4 Stop valve shower 5 Fill in 6 Thermostat mixing valve 7 Overflow 8 Hydro jets 9 Up down button 10 Drain 11 Adjustable foot 12 Hydro bu...

Page 8: ...7 Components Disinfection with stop valve Fittings Hydro system...

Page 9: ...e Note For safety reasons the temperature is mechanically limited to 38 C standard mixer To override this security appliance press the red button on the knob Now it s possible to reach a higher temper...

Page 10: ...tivated an accurately meas ured amount of disinfectant is directed into the wand for optimum disinfection and cleaning Press and hold the black trigger directly on the red shower wand and spray all in...

Page 11: ...nt to stop the therapy you have to press the button again Fill in On To activate the fill in press the chromed button which is labelled with On Water will start coming out of the fill in spout On a pr...

Page 12: ...es the above mentioned restrictions Always clean the plastic parts wet or damp but never dry We recommend using our approved shampoo and bath additive products with your GK bath Other products may cau...

Page 13: ...wer wands and clean them if neces sary When servicing the bathtub the unit must be unplugged and switched off Service At least once a year a service or safety inspection in accordance with Medical Dev...

Page 14: ...egulations The bathtub complies with the essential requirements pursuant to Appendix I of the Medical Device Directive 93 42 EEC In accordance with the classification regulations pursuant to Ap pendix...

Page 15: ...on class I connection of protective earth conductor necessary or II see identification plate on the bathtub Degree of protection IPX4 Protection type Type B Lifting weight 500 kg Lifting height 250 mm...

Page 16: ...Pa to 1060 hPa Transport and Storage Ambient temperature range 20 C to 60 C Relative humidity range 80 not condensed at 20 C Atmospheric pressure range 500 hPa to 1060 hPa Recycling Package Wood and c...

Page 17: ...Notes 16 Notes...

Page 18: ...Notes 17 Notes...

Page 19: ...Notes 18 Notes...

Page 20: ...ents All product related information is not subject to change for technical production or other reasons Georg Kr mer Ges m b H Co KG A 4061 Pasching Neuhauserweg 14 Tel 43 0 7229 64450 Fax 43 0 7229 7...