20
20
EN
EN
USER’S GUIDE
IMP20
5. PARAMETER SETTING
5.1 ESSENTIAL PARAMETER SETTING
Press
SET
and input
18
, then press
F1 1
Zero function range setup
, choose
0
means 0%
,
1
means 2%,
2
means 20%,
3
means 50%,
4
means 100%, then press
to confirm.
F2 1
Zero function range of power on
, choose
0
means 0%,
1
means 4%,
2
means 50%,
3
means 100%, then press
to confirm.
F3 05
Zero-tracking range
, choose
5
means 0.5e,
10
means 1e, max is
20
means 2e,
then press
to confirm.
F4 4
Baud rate of wireless
, choose
1
means 600bps,
2
means 1200bps,
3
means 4800bps,
4
means 9600bps, then press
to confirm.
F5 4
Baud rate of RS232
, choose
1
means 600bps,
2
means 1200bps,
3
means 4800bps,
4
means
9600bps, then press
to confirm.
5.2 SETUP PRINTING PARAMETER
Press key
SET
, input
97
and press
P1 0
Setup print out whether back to
0
or not, choose
0
means back to 0 and print,
1
for not back
to 0 and print, then press
to confirm.
P 2 1
Setup copies of printing bill, choose
1
means 1copy,
2
means 2 copies,
3
means 3 copies, then
press
to confirm.
P 3 0
Setup time of auto printing after stable light on, choose
0
means the time after stable,
1
means
1s after stable,
2
means 2s and the longs time is 3s.