USER’S MANUAL
ACM
Key (Order)
Step Explanation
LCD Screen Display
Short press
key
1. Machine Tare
0.00g
Press and Hold key for 5
seconds.
2. Display Span CAL signal Flash 200g signal
200.00g
Release key, Press and
Hold key for 5 seconds.
3. Display Linearity CAL signal Flash 100g
signal
100.00g
4. Put required weight mass on pan Display
100g after 5 seconds.
100.00g
5. Take weight mass away (Lenearity calibration
finished)
0.00g
Note: The grey color words explain the signal’s meaning which flash on the window.
IV
FOURTH PART: OPERATE APPLICATION
APPLICATION SETTING
COUNTING
PURPOSE:
Use this function can calculate the wuantity with tota lweight divide by signal weight.
INSTANCE:
(With know the sample’s quantity but unknown the unit weight)
Basic Weighing (The machine warm up)
Key (Order)
Step Explanation
LCD Screen Display
Short Press key
1. Enter into counting function and display flash
the number of quantity.
Short press
key to cycle the sample quantity setting:
10, 20, 50, 100, 150, 200, 250, 500, 1000pcs.
2. Put any unknown numbers of pieces on pan.
(Example: set 20pcs)
20 pcs
Short Press Key
3. The machine start to calculate and display
quantity: 20pcs
20 pcs
4. Take sample away from pan
0 pcs
5. Put any unknown numbers of pieces on pan and
will display a count. (Example: 100pcs)
100
pcs
Short Press Key
6. Exit the counting function
0.00 g