UM bilancia analitica BL224 ENG rev.0 13.05.2020
25
Check Weighing
This function allows to control whether a given sample's weight falls within a certain range of values, or if it reaches a
certain target value.
Check weighing
Procedure
1.
Make sure you have selected the "
F CHw
" function in the user menu.
2.
Adjust the scale by pressing the <
TARE
> key.
3.
Press the <
UNIT
> key until "
CW
" appears on the display
Now the balance needs to know the tolerance interval and the target value
Tolerance interval and the target value settings
1.
Press and hold the
▲
key to enter the setup menu
2.
Use the
▲
button to scroll through the
TARGET, HI, LOW
items
3.
Press the <
TARE
> key on the parameter you want to change
4.
Use the
▲
and
▶
keys to set the desired number.
5.
Press the <
TARE
> key to confirm
6.
Press and hold <
TARE
> to save the settings and return to size.
7.
7Place the sample on the plate, the display will show the result in
accordance with the inserted interval and target settings.
•
If the
LL
sign appears, the sample weight is below the minimum limit of the
set interval.
•
If
HH
appears, the sample weight is greater than the maximum limit of the
set interval.
•
In addition to LL or HH, if the sample weight is outside the set range, the
balance
will sound
.
•
To display the weight of the sample when the scale shows LL or HH, press the
<
CAL
> key