![Gima TEMP SITTER User Manual Download Page 2](http://html1.mh-extra.com/html/gima/temp-sitter/temp-sitter_user-manual_2228222002.webp)
• Adults should help children wear and operate thr device.
• Please keep the arm straight for at least 5-8 minutes to obtain exact
temperature.
• After measurement, replace the sensor in the packaging to avoid battery
consumption, as the device is always on.
Note:
Different measurement formats and methods can lead to different
temperature values.
Maintenance and cleaning
1 Clean the surface of the device before and after use with 75% concentrate
medical alcohol;
2 Stop using the device when the battery indicates low battery, otherwise
the device will not operate correctly;
3 The product has a useful life of 3 years. Disposal of the device and its battery
must be carried out at the end of its useful life in accordance with national
laws;
4 The probe must be cleaned and disinfected, but never sterilised. Do not
sterilise the device in an autoclave or otherwise
• If the probe shows signs of damage or deterioration, stop using the device
immediately.
• The use of detergents or disinfectants other than alcohol may damage
the device and reduce its usable life, or cause a safety risk.
• The electrical screen and the list of components is only supplied to
assistance centres or qualified personnel certified by the manufacturer.
• Please refer to the product label for the manufacturing date.
Product specifications
Possibili Problemi e Soluzioni
Guandong Biolight Meditech Co., Ltd.
No. 2 Innovation First Road, Technical Innovation Coast
Hi-Tech Zone, Zhuhai, China
Made in China
25548 / WT1
M-25548-GB-Rev
.0.09.17
Product specifications
Definition of symbols
Manufacturer EMS Guide and Declaration
Guarantee Certificate
If you require technical advice or encounter malfunctions with your device
(excluding human error or defects cause by accidents) during the guarantee
period, please contact the assistance centre.
NOTE:
the same registration and password cannot be shared between
multiple devices; we recommend creating a new account for each mobile
device.
Production Model
WT1
Size
51mm×31.8mm×8.5mm
Weight
9.6g
Measurement specifications
Measurement range
25.00°C~45.00°C (77.00°F~113.00°F)
Measurement accuracy
±0.05°C (35.00°C~38.50°C) \
±0.09°F (95.00°F~101.30°F)
±0.1°C (25.00°C~34.99°C and 38.51°C~45.00°C) \
±0.18°F (77.00°F~94.99°F) and (101.31°F~113.00°F)
Measurement units for temperature “°C” or “°F”
Update frequency
Every 3s
Illuminated Bluetooth indicator
Not connected: flashes twice continuously.
Connected: flashes once.
App software platform
iOS System
iPhone 4s+, iPad3+, iPad mini
Android System (4.3+)
Mobile telephone, Tablet
Battery
Type
CR2032, 3V/225mAH button battery
Operating time
≥365 days (8 hours per day)
Environmental specifications
Temperature
In use: 5°C~+40°C;
Transportation and storage: -20°C~+55°C
Humidity
In use: 25%~95% (without condensation);
Transportation and storage: 25%~95%
(without condensation)
Atmospheric pressure
In use: 700hPa~1060hPa;
Transportation and storage: 500hPa~1060hPa
Problem
Possible Cause
Solution
No sound warning
The sound system is off
Cannot set the sound system
correctly
Unable to establish a
connection between
the device and the
receiver
1. Bluetooth on the receiver
is off
2. The Android receiver
version is not 4.3 or
greater
3. The Bluetooth version is
not 4.0
4. Out of range
5. Incorrectly positioned
1. Switch Bluetooth on
2. Update the Android system to
version 4.3 or greater
3. Use a receiver device with
Bluetooth 4.0
4. Switch off the receiver
5. Change the receiver position
Cannot switch the
device on
1. The battery may be low
or power missing
2. The device itself may be
damaged
1. Restart the device
2. Contact customer services
Cannot see data
correctly regarding
temperature
1. Device battery is flat
1. Restart the device
2. Contact customer services
Number
Components
WT1
1
Adhesive tape
One
2
User manual
One
Product name
Model name
Product serial
number
Guarantee Period
User Name
User address
Contacts
Telephone contact
Manufacturer Guide and Declaration - Electromagnetic Emissions
The device is designed for use in electromagnetic areas as specified below.
The customer or user must ensure that the product is used in such areas.
Emissions test
Compliance
Guide - electromagnetic
environment
RF Emissions
CISPR 11
Group 1
The device uses RF energy only
for its internal functions. Its RF
emissions are therefore very low,
and it is improbably that they
may cause interference with other
electronic devices nearby.
RF Emissions
CISPR 11
Class B
The device is suitable for use in all
installations, including domestic
installations and those connected
directly to a low-voltage public
power supplier that supplies
buildings for domestic use.
Harmonised emissions
IEC 61000-3-2
Not applicable (Device
operates on batteries)
Current fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable (Device
operates on batteries)
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537, Hamburg, Germany
0123
IP27
Simbolo
Nota
Keep away from sunlight
Keep in a cool, dry place
Read instructions carefully
Products should not be disposed of with household waste
Part applied of CF type without defibrillation protection
IP27
Casing protection class
Refer to this user manual
SN
Serial number
Product complies with European Directive no. 98/79/EC
Manufacturer
Authorized representative in the European community
Product code
Lot number