WARNINGS & PRECAUTIONS DURING THE NIBP MESURAMENT
This device should not be used when oscillometric pulses may be altered by other
devices or techniques such as External Counterpulsation (ECP) or Intra Aortic Balloon
Pump Counterpulsation.
DO NOT
USE THE
NIBP MODULE
FOR ANY PURPOSE OTHER THAN SPECIFIED IN THIS MANUAL
WITHOUT WRITTEN CONSENT AND APPROVAL FROM
ELPRO
S
.
R
.
L
.
DO NOT
USE IN THE
PRESENCE OF FLAMMABLE GASEOUS ANESTHESIA AGENTS
BECAUSE
OF FLAME HAZARD
.
DO NOT
ATTACH THE CUFF TO A LIMB BEING USED FOR IV INFUSIONS AS THE CUFF INFLATION
CAN BLOCK THE INFUSION
,
POTENTIALLY CAUSING HARM TO THE PATIENT
.
SUBSTITUTION OF A COMPONENT DIFFERENT FROM THAT SUPPLIED MAY RESULT IN
MEASUREMENT ERROR
.
REPAIRS SHOULD BE UNDERTAKEN ONLY BY PERSONNEL
TRAINED OR AUTHORIZED BY
ELPRO
S
.
R
.
L
.
Accuracy of any blood pressure measurement may be affected by the position of the
subject, his or her physical condition and use outside of the operating instructions
detailed in this manual. Interpretation of blood pressure measurements should be made
only by a physician or trained medical staff.
The RESCUE LIFE NIBP module is designed to work with SunTech
®
cuffs and hoses. The
use of cuffs and hoses not supplied by SunTech
®
may compromise performance and
accuracy.
If the blood pressure cuff is on the same limb as a pulse oximeter probe, the oxygen
saturation results will be altered when the cuff occludes the brachial artery.
To obtain accurate blood pressure readings, the cuff must be the correct size, and also
be correctly fitted to the patient.
Incorrect size or incorrect fitting may result in
incorrect readings.
When a cuff is going to be positioned on a patient for an extended length of time, be
sure to occasionally check the limb for proper circulation. The module may not operate