Gima LS808-BS Manual Download Page 5

33

•  Please check that operation of the device does not result in prolonged impairment of patient 

blood circulation.

•  When measurement, please avoid compression or restriction of the connection tubing.

•  The device cannot be used with HF surgical equipment at the same time.

•  The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clin-

ically investigated according to the requirements of ISO 81060-2:2013.

•  To  verify  the  calibration  of  the AUTOMATED  SPHYGMOMANOMETER,  please  contact  the 

manufacturer.

•  This device is contraindicated for any female who may be suspected of, or is pregnant.

  Besides providing inaccurate readings, the effects of this device on the fetus are unknown.

•  Too frequent and consecutive measurements could cause disturbances in blood circulation and 

injuries.

•  This unit is not suitable for continuous monitoring during medical emergencies or operations. 

Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and even purple due 

to a lack of blood.

•  When not in use, store the device with the adapter in a dry room and protect it against extreme 

moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage 

case.

•  This device may be used only for the purpose described in this booklet. The manufacturer 

cannot be held liable for damage caused by incorrect application.

*  This device comprises sensitive components and must be treated with caution. Observe the 

storage and operating conditions described in this booklet.

•  The maximum temperature that the applied part can be achieved is 42.8℃ while the environ-

mental temperature is 40℃.

•  The equipment is not AP/APG equipment and not suitable for use in the presence of a flamma-

ble anesthetic mixture with air of with oxygen or nitrous oxide.

•  Warning: No servicing/maintenance while the ME equipment is in use.

•  The patient is an intended operator.

•  The patient can measure,transmit data and charge power under normal circumstances and 

maintain the device and its accessories according to the user manual.

•  To avoid measurement errors, please avoid the condition of strong electromagnetic field radiat-

ed interference signal or electrical fast transient/burst signal.

•  The blood pressure monitor, its adaptor, and the cuff are suitable for use within the patient 

environment. If you are allergic to polyester, nylon or plastic, please don’t use this device.

•  During use, the patient will be in contact with the cuff. The materials of the cuff have been tested 

and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010.

  It will not cause any potential sensation or irritation reaction.

•  Adaptor is specified as a part of ME EQUIPMENT.

•  If you experience discomfort during a measurement, such as pain in the arm or other com-

plaints, press any button to release the air immediately from the cuff. Loosen the cuff and 

remove it from your arm.

•  If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the 

cuff not deflate when pressure reaches 40 kPa (300 mmHg), detach the cuff from the arm and 

press any button to stop inflation.

•  Before use, make sure the device functions safely and is in proper working condition.

  Check the device, do not use the device if it is damaged in any way. The continuous use of a 

damaged unit may cause injury, improper results, or serious danger.

•  Do not wash the cuff in a washing machine or dishwasher!

  The service life of the cuff may vary by the frequency of washing, skin condition, and storage 

state. The typical service life is 10000 times.

ENGLISH

Summary of Contents for LS808-BS

Page 1: ...MEDICAL PRODUCTS Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District Zhongshan 528437 Guangdong China Made in China 32916 LS808 BS MDSS Medical Device S...

Page 2: ...DATA MANAGEMENT 44 Recall the Records 44 Delete the Records 46 INFORMATION FOR USER 47 Tips for Measurement 47 Maintenance 48 ABOUT BLOOD PRESSURE 49 What are systolic pressure and diastolic pressure...

Page 3: ...ntended for adult indoor use only Contraindications 1 The device should not be used by any person who may be suspected of or is pregnant 2 The device is not suitable for use on patients with implanted...

Page 4: ...r blood pressure When the device is used to measure patients who have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation the best result may occur with deviation P...

Page 5: ...use in the presence of a flamma ble anesthetic mixture with air of with oxygen or nitrous oxide Warning No servicing maintenance while the ME equipment is in use The patient is an intended operator Th...

Page 6: ...ort to Gima if any unexpected operation or events occur Keep the unit out of reach of infants young children or pets to avoid inhalation or swallowing of small parts It is dangerous or even fatal Be c...

Page 7: ...inks when the blue tooth is working Average value The average value of the latest three groups bood pressure value Memory Query Indicate it is in the memory mode and which group of memory it is User 1...

Page 8: ...sure Monitor 2 AC Adaptor LS808 BS Model BLJ06L060100P V 3 User Manual 4 Cuff 22cm 42cm Type BF Applied Part Please use Gima authorized cuff The size of the actual cuff please refer to the label on th...

Page 9: ...d for about 20 days Storge and use the blood pressure monitor at the cool dry and ventilated environment Avoid to approach to the fire and the heat source or it will cause the battery explode Only can...

Page 10: ...Setting the Time Date and Unit To ensure the stored measurement result has correct time record please set time and unit before device is used Before use switch the button to the ON side to turn on th...

Page 11: ...39 3 Press User 1 button again to confirm YEAR Then the numeral representing MONTH blinks 4 Repeat step 2 and 3 to confirm MONTH and DAY 5 Repeat step 2 and 3 to confirm HOUR and MINUTE ENGLISH...

Page 12: ...ure Monitor with Your Device 1 Turn on Bluetooth and the app Make sure both are ON when pair up is proceeding 2 In the search device interface please select the device to pair up 3 When the monitor is...

Page 13: ...will shut off after Pair up process is complete Bluetooth Module No LS51802 RF Frequency Range 2402 MHz to 2480 MHz Output Power Range 0 dBm Supply Voltage 1 8 3 6 V Transmitting Distance 10 meters Li...

Page 14: ...n in your arm use the other one 2 Roll or push up your sleeve to expose the skin 3 Apply the cuff to your arm with your palm facing up 4 Position the edge of the cuff about 2cm 3cm from elbow 5 Fasten...

Page 15: ...sure data Take User 1 for example Note Select the same user on your app and BPM to take the measurement or the measurement data won t be transmitted to the app 1 When the monitor is off press the User...

Page 16: ...DATA MANAGEMENT Recall the Records 1 When the monitor is off please press the Query button to show the average value of the latest three records If the records are less than three groups it will displ...

Page 17: ...is January 2nd is 8 00 CAUTION The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and...

Page 18: ...60 is dropped from the list Delete the Records 1 Hold pressing Query button for 3 seconds when the monitor is in the memory recall mode the dEL ALL User ID will flash on the display 2 Press Query to...

Page 19: ...taken in the following circumstances Within 1 hour after dinner or drinking Immediate measurement after tea coffee smoking Within 20 minutes after taking a bath When talking or moving your fingers In...

Page 20: ...t in a dry place and avoid the sunshine Avoid immersing it in the water Clean it with a dry cloth in case Avoid intense shaking and collisions Avoid dusty environment and unstable temperature surround...

Page 21: ...ysician can tell whether your blood pressure value has reached a dangerous point Irregular Heartbeat Detector An irregular heartbeat is detected when a heartbeat rhythm varies while the device is meas...

Page 22: ...the pressure will vary more 3 Wait at least 3 minutes for another measurement Why do I get a different blood pressure at home compared to the hospital The blood pressure is different even throughout...

Page 23: ...ust the cuff and relax for a moment and then measure again E 10 or E 11 shows The monitor detected motion talking or the pulse is too poor while measuring Relax for a moment and then measure again E 2...

Page 24: ...ange of 700 hPa to 1060 hPa Storage transportation condition Temperature 20 C to 60 C A relative humidity range of 93 non condensing at a water vapour pressure up to 50hPa Measurement perimeter of the...

Page 25: ...V Output 100 240V 50 60Hz 0 2Amax Input 6V 1000mA Contact Information Manufactured by Guangdong Transtek Medical Electronics Co Ltd Company Guangdong Transtek Medical Electronics Co Ltd Address Zone...

Page 26: ...al standard Electromagnetic disturbances Requirements and tests Performance requirements EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 Requirements and test methods for non automated measu...

Page 27: ...l antennas should be used no closer than 30 cm 12 inches to any part of the equip ment LS808 BS including cables specified by the manufacturer Otherwise degradation of the performance of this equipmen...

Page 28: ...rrup tions and voltage variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 250 300 cycle...

Page 29: ...0 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2...

Reviews: