4.6 Troubleshooting
When there is a fault with the urine analyzer or some functions can not be
achieved caused by users’ improper operation, the urine analyzer will displays
error message, all error messages are as follows:
Error message
Abnormal system
information
Abnormal Motor
Abnormal light
source
Test paper is placed
wrong
Can’t power on
Solution
The memory has problems, the analyzer can not properly
read the system parameters, if it still displays abnormal message
after restarting please contact the dealer.
1. Motor rotation is abnormal, check whether there is debris on
the paper tray, causing the motor stuck.
2. Low battery, restart after connecting the AC adapter.
1. The light source is strong, weak or damaged.
2. The calibration comparator block is contaminated,
please clean it.
Please check whether the head of the test strip has been placed
to the inner end of the tray flute, even the strip has not been
placed, if it is, please correct it within 10 seconds and press
OK button.
Low battery, restart after connecting the AC adapter, if still can’t
power on, please contact the dealer.