NOTE:
If the calendar clock time has not been set, the device will enter the Time
Settings screen while it powers up.
The device will enter standby status automatically if there is no key-press and
no measuring signal in 20 seconds.
That is the only way to turn it off (i.e. at standby status).
3.1 Taking Measurement
1.
Insert the SpO
2
probe connector into the SpO
2
probe receptacle on the de-
vice properly.
SpO
2
probe connector
Figure 5 SpO
2
Probe Connection
2. Fix the oximeter on your left wrist (Figure 6);
Fingertip sign
Figure 6 Placement Illustration
3. Then, hold the probe with its opening towards your index finger. The probe
should be oriented in such a way that the sensor side with a finger tip icon is
positioned on the top (Figure 6).
4. Next, insert your index finger into the probe until the fingernail tip rests
against the stop at the end of the probe (Figure 7). Adjust the finger to be
placed evenly on the middle base of the sensor (make sure the finger is in
the right position). If the index finger cannot be positioned correctly, or is not
available, other finger can be used.