Ultrasonic cleaning is strictly prohibited.

2. Disinfection

Soaking in solutions or Thermo chemically in a washer Sterilizer up to 65°C maximum may perform disinfection. 

Manufacturers instruction regarding duration and concentration of solutions should be strictly adhered. After Disin-

fection, rinse throughly in sterile water and dry with a lint free clean cloth.

3. Cold Soak Solution

To achieve a high-level disinfection, Cidex®OPA or 2.4% Glutaraldehyde solution may be used according to manu-

facturer’s instructions. Dry with lint free, clean cloth or filtered pressurized air. Reassemble all parts, load handle with 

batteries and test the system for proper function.

If not functional, review battery / lamp testing instructions below.

Do not immerse Blades in Bleach, Betadine or Potassium Hydroxide solutions.

Doing so will several damage instruments also avoid metal to metal contact after soaking, the blades should 

be rinsed under sterile water to remove chemical residues and dry with lint free clean cloth or filtered pres-

surized air.

4. Sterilization

Before performing any of the procedures described below, the blade should be cleaned as described in the cleaning 

Procedure.

5. Gas Sterilization

Gas sterilization by Ethylene oxide up to a maximum temperature of 65°C and 8 p.s.i, may be performed, which is 

preferred especially if sterilization is to be performed regularly.

6. Steam Sterilization

Steam Sterilization can also be performed. Insert device in appropriate autoclave pouch.

(A) GRAVITY DISPLACEMENT STEAM

(B) PRE-VACUUM STEAM

Temperature

121°C (250°F) 

134°C (273°F)

Cycle Time

30 Minutes

5 Minutes

Dry Time

15 Minutes

20 Minutes

Note: Do not exceed temperature of 135°C and pressure of 28 p.s.i

Flash autoclaving and hot air sterilization should be avoided as these processes will damage the instrument.

Steris Amsco V-Pro

Laryngoscope Blades and Handles are compatible with:

Amsco V-Pro 1 Low Temperature Sterilization System

Amsco V-Pro 1 Plus Low Temperature Sterilization System

Amsco V-Pro 1 Pro Max Temperature Sterilization System

Sterrad

Larynogoscope Blades and Handles are compatible with:

Sterrad 100nx System (Standard and Express Cycle)

Sterrad nx System (Standard Cycle)

Sterrad 1005 and 200 System (Short Cycle Outside US)

Sterrad 50 System

HANDLES

Cleaning/Sterilization:

Battery Handle withstand the same cold soak solution and autoclave ranges outlined in the blade section. However, 

the batteries and lamp must be removed prior to disinfection/sterilization.

Battery Handles can withstand exposure to the ethlene oxide. Lamp may be cleaned with cotton ball dampened in 

alcohol (IPA). The main handle and a cap will also withstand and exposure to ethylene oxide.