Care and Maintenance

1. Cleaning Procedure:

Immediately after use, the Killian Nasal Speculum should be rinsed under cool running tap water until all 

visible soil is removed. Ensure that all hard-to-reach areas are flushed with the running tap water.

Immerse sealed Killian Nasal Speculum system in a presoak enzymatic cleaner solution, which was 

prepared in accordance to manufacturer's recommendations for a minimum of two minutes.

Remove device from enzymatic cleaner solution and rinse with lukewarm running tap water for a mini-

mum of one minute to remove all residues and visible soils. Then, immerse device in enzymatic deter-

gent. Brush thoroughly using a soft-bristle brush, while ensuring that all hard to reach are reached visible 

soils / residue removed. Dry with lint free, clean cloth or filtered pressurized air. Follow with HIGHLEVEL 

DISINFECTION or STEAM STERILIZATION PROCEDURE.

2. Disinfection:

Soaking in solutions or Thermo chemically in a washer Sterilizer up to 93°C maximum may perform dis-

infection. Manufacturer's instruction regarding duration and concentration of solutions should be strictly 

adhered. After Disinfection, rinse thoroughly in sterile water and dry with a clean lint free cloth.

3. Cold Soak Solution:

To achieve a high-level disinfection, Cidex®OPA or 2.4% Glutaraldehyde solution may be used according 

to manufacturer's instructions. Dry with lint free, clean cloth or filtered pressurized air.

Do not immerse Killian Nasal Speculum in Bleach, Betadine or Potassium Hydroxide solutions. Doing 

so will several damage instruments also avoid metal to metal contact after soaking, the Killian Nasal 

Speculum should be rinsed under sterile water to remove chemical residues and dry with lint free clean 

cloth or filtered pressurized air.

4. Sterilization:

Before performing any of the procedures described below, the Killian Nasal Speculum should be cleaned 

as described in the cleaning Procedure.

5. Gas Sterilization:

Gas sterilization by Ethylene oxide up to a maximum temperature of 65°C and 8 psi may be performed, 

which is preferred especially if sterilization is to be performed regularly.

6. Steam Sterilization:

Steam Sterilization can also be performed. Insert device in appropriate autoclave pouch.

Note:

Do not exceed temperature of 135°C and pressure of 28 psi Flash autoclaving and hot air sterilization 

should be avoided as these processes will damage the instrument.

Warning:

a) Read the label before opening packing.

b) Handle it with care.

c) Only trained professional should use it.

d) Use protective gloves and wash the hand with anti-septic solution or Anti -bacterial soap before oper-

ating any medical operation.

e) Do not leave unattended around children.

f)  Don't keep it in severe conditions.

g) Store it at specified temperature.

(A) GRAVITY

DISPLACEMENT STEAM

(B) PRE-VACUUM STEAM

Temperature 

Cycle time 

Dry time

121°C (250°F)

30 Minutes

15 Minutes

134°C (270°F)

5 Minutes

20 Minutes

ENGLISH

Summary of Contents for 26777

Page 1: ...FIBER OPTIC LIGHT M26777 M Rev 0 01 19 PROFESSIONAL MEDICAL PRODUCTS Manuale d uso User manual ATTENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual before using the product Gima S p A Via Marconi 1 20060 Gessate MI Italy Made in Pakistan 26777 ...

Page 2: ... so that when the handles are squeezed together the blades spread laterally allowing examination The outer part of Killian Nasal Speculum is made up of two main blades and is merged in each other with a hinge It gives shape to the Killian Nasal Speculum One of the main part is Light Guide Tube which is the source of light for viewing the internal parts of nose nostrils No substance like Latex pote...

Page 3: ...ructions Dry with lint free clean cloth or filtered pressurized air Do not immerse Killian Nasal Speculum in Bleach Betadine or Potassium Hydroxide solutions Doing so will several damage instruments also avoid metal to metal contact after soaking the Killian Nasal Speculum should be rinsed under sterile water to remove chemical residues and dry with lint free clean cloth or filtered pressurized ai...

Page 4: ...iginal spare parts defects caused by negligence or incorrect use GIMA cannot be held responsible for malfunctioning on electronic devices or software due to outside agents such as voltage changes electro magnetic fields radio interferences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected produc...

Reviews: