![Gima 2219B Quick Start Manual Download Page 5](http://html.mh-extra.com/html/gima/2219b/2219b_quick-start-manual_2227839005.webp)
9
9 MAINTENANCE
1. After every treatment, please only use neutral solution to clear the the dust, stains on leather, metal part or plastic cover. Disinfecting solution
like alcohol, acid or alkaline solution are forbidden. Th is is to prevent the lifetime of leather and plastic cover from being shorten.
2. Please lubricate the moving parts regularly (once a year).
10 CIRCUIT SYSTEM CONNECTION DIAGRAM
11 TROUBLESHOOTING
If problem is still not solved after above operation, please contact maintenance center or supplier to solve it.
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing
it to a specific recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
ENGLISH
Power
Hand controller
Foot controller
(Option)
Backrest motor
Whole bed lifting motor
Control Box
THE FAULT PHENOMENON
REASON
SOLUTION
Can't move up or down
normally
1. The power cord is unconnected or not plugged well
1. Connect the power cord or plug it well
2. Poor quality of the hand controller or foot controller
2. Replace the hand controller or foot controller
3. Poor quality of the control box
3. Please contact maintenance center or supplier
for troubleshooting.
4. Repture of motor tail
5. Motor was damaged
Abnormal Sound
1. The rotating parts are short of lubricating oil
1. Add lubricating oil
2. Rotating positions (such as hinge, POM sets) are
short of lubricating oil
2. Add lubricating oil
3. Motor quality problems
3. Replace the motor