31
1
×
12]/[Rhythm 12]/[Rhythm
10]/[Rhythm 8]/ [Rhythm 6]
Lead Gain
[Smart]/[Current]
The option selected will be used as printing
gain mode. "Smart" means the system will
adjust gain automatically to fit paper height;
"Current" means it will use screen waveform
gain as that of printing.
Auto Strip
[2.5 sec]/[3 sec]/[4 sec]/[5
sec]/[6 sec]/[8 sec]/[10
sec]/[15 sec]/[20 sec]/[25 sec]
The system takes the selected option as the
print time length of each strip.
Rhythm
Strip
[10 sec]/[15 sec]/[20 sec]/[25
sec]/[30 sec]
When “Print Mode” is set to “Rhythm 12”
“Rhythm 10” “Rhythm 8” or “Rhythm 6”,
the system takes the selected option as the
print time length of each waveform.
Average
QRS
[3x4 + Mark]/[3
×
4]/[Not
Print]
When “Print Mode” is set to “Auto” or
“Rhythm”, the system uses the selected
format to print the average QRS waveform.
Auto-Diagn
osis
[All]/[Only Data]/[Only
Conclusion]/[Not Print]
The diagnosis contains data and conclusion
two parts, which can be chosen by user as
demand.
Period
[Off]/[per 1 min]/[per 2 min]
/[per 3 min] /[per 5 min]/[per
10 min]/[per 20 min]/[per 30
min]/[per 60 min]
During the ECG acquisition process, the
system will automatically activate the
printing operation according to the selected
time interval. When the printing mode is
manual mode, the printing will output “Auto
12
×
1” format, otherwise, it will output
according to the current setting mode.
Print
Device
[Inside]/[Outside A4]
Choose to print out the ECG waveform by
thermal printing system or USB external
printer
Print Depth
[1]/[2]/[3]/[4]
Set the waveform depth according to
necessary.
Timing
Marker
ON/OFF
Set to display the time mark on the print
paper or not
Arrhythmia
ON/OFF
Set to turn on or off the arrhythmia analysis
Note 1: The settings of auto strip, rhythm strip, average QRS, auto diagnosis and
period are only optional in auto mode and rhythm mode.
Note 2: If print time length is less than 8s, the sampling and analysis time will be 8s;
if the print time length is equal to or greater than 8s, the sampling and analysis time keeps
the same with print time.