TROUBLESHOOTING
|
PIPETMAN® CLASSIC
Leak Test
This test may be performed at any time to check
that the pipette does not leak, especially after
performing a maintenance or decontamination
procedure. If a pipette fails this test, replace
the O-ring and seal. After making sure that the
pipette is correctly reassembled, repeat this test.
P2 to P200 Models
1.
Fit a PIPETMAN® DIAMOND Tip.
2. Set the pipette to the maximum volume,
given in the specifications and pre-rinse.
3. Aspirate the set volume from a beaker of
distilled water.
4. Maintain the pipette in the vertical position
and wait for 20 seconds.
5. If a water droplet appears at the end of the
tip, there is a leak.
6. If you do not see a droplet, re-immerse the
tip below the surface of water.
7. The water level inside the tip should remain
constant. If the level goes down, there is a
leak.
P1000 to P10mL Models
1.
Fit a PIPETMAN DIAMOND Tip.
2. Set the pipette to the maximum volume
given in the specifications.
3. Aspirate the set volume from a beaker of
distilled water.
4. Maintain the pipette in the vertical position
and wait for 20 seconds.
5. If a water droplet appears at the end of the
tip, there is a leak.