Leak Test

This test may be performed at any time to check that the pipette does not leak, especially after 

performing a maintenance or decontamination procedure. If a pipette fails this test, replace the 

O-ring and seal. After making sure that the pipette is correctly reassembled, repeat this test.

For the P2 to P200 single channel models:

1. 

Fit a PIPETMAN DIAMOND Tip.

2.  Set the pipette to the maximum volume given in the specifications and pre-wet.
3.  Aspirate the set volume from a beaker of distilled water.
4.  Maintain the pipette in the vertical position and wait for 20 seconds:

 

If a water droplet appears at the end of the tip, there is a leak.

 

If you see no droplets, re-immerse the tip below the surface of water. The water level inside the 

tip should remain constant; if the level goes down, there is a leak.

For the P1000, P5000, and P10mL single channel models:

1. 

Fit a PIPETMAN DIAMOND tip.

2.  Set the pipette to the maximum volume given in the specifications.
3.  Aspirate the set volume from a beaker of distilled water.
4.  Maintain the pipette in the vertical position and wait for 20 seconds:

 

If a water droplet appears at the end of the tip, there is a leak.

For the multichannel models:

1. 

Fit the PIPETMAN DIAMOND tip.

2.  Set the pipette to the maximum volume given in the specifications, and pre-wet the tips.
3.  Aspirate the set volume from a reagent reservoir of distilled 

water.

4.  Maintain the pipette in the vertical position and wait for 20 

seconds:

 

If a water droplet appears at the end of the tip, there is a 

leak.

 

If you see no droplets, for volumes below 200 µL, 

re-immerse the tip below the surface of water.

 

The water level inside the tip should remain constant; 

there is a leak if the level goes down.

no leak

Figure 10

MyPIPETMAN® multichannel leak test

Chapter 6

CLEANING AND DECONTAMINATION

MyPIPETMAN pipettes are fully autoclavable without disassembly for maximum convenience as well as 

protection from contamination.

NOTICE

You may refer to the decontamination procedure available on

 

www.gilson.com

.

Liquid must never enter any pipette’s upper part (handle).

Cleaning for Single Channel Models

The pipette must be cleaned, as described below, before it is decontaminated. Soap solution is 

recommended for cleaning MyPIPETMAN.

External

1. 

Remove the tip ejector (refer to 

Changing the Tip Ejector for Single Channel Models

).

2.  Wipe the entire pipette and the tip ejector with a soft-cloth or lint-free tissue soaked with soap 

solution to remove all dirty spots or marks. If the pipette is very dirty, a brush with soft, plastic 

bristles may be used.

3.  Wipe the entire pipette and the tip ejector with a soft cloth or lint-free tissue soaked with distilled 

water.

4.  Refit the tip ejector and allow the pipette to dry.

Summary of Contents for MyPipetman

Page 1: ...User s Guide EN A U TOCLAVABL E FULLY...

Page 2: ...r Good Pipetting 10 ACCESSORIES 11 MyPIPETMAN Accessories for Extra Personalization 11 TROUBLESHOOTING 14 Troubleshooting Table 14 Leak Test 15 CLEANING AND DECONTAMINATION 15 Cleaning for Single Chan...

Page 3: ...tte Colors Special Text Printing Special Logo Printing Ordering Information Minimum Order Quantity Available Countries 1 Worldwide Possible on Request Subject to Specific Conditions 150 Units Mix of D...

Page 4: ...and MyPIPETMAN Enterprise serial number type A A 50001 MyPIPETMAN serial number type MY 524 A A Figure 2 Serial number identification example Year Month Production number Model Random Number Month Ye...

Page 5: ...6 0 3 MyPIPETMAN P200 D200 20 100 200 0 50 0 80 1 60 0 20 0 25 0 30 2 5 0 8 0 8 1 0 0 25 0 15 1 6 1 6 1 6 0 6 0 6 0 6 8 0 1 6 0 8 3 0 0 6 0 3 MyPIPETMAN P1000 D1000 100 500 1000 3 0 4 0 8 0 0 6 1 0 1...

Page 6: ...00 L 0 2 L MyPIPETMAN P8x300 L 0 2 L MyPIPETMAN P12x300 L 0 2 L 1 5 P2 1 25 L 2 0 7 5 P10 7 5 L 1 2 5 P20 12 5 L P100 0 7 5 75 L P200 1 2 5 125 L P1000 0 7 5 0 75 mL P5000 1 2 5 1 25 mL P10mL 0 7 5 7...

Page 7: ...r extension refer to Figure 7 1 Slide the extension over the tip holder 2 Push the extension firmly onto the end of the tip ejector until it clicks into place To remove a tip ejector extension 1 Gentl...

Page 8: ...accuracy and precision by creating an efficient barrier to aerosols or corrosive vapors such as acids which could damage mechanical parts Insert the filter directly in the pipette tip holder Change th...

Page 9: ...wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoothly to the first stop 3 Wait for at least a second then press the push button to the second stop to expel...

Page 10: ...the tip For volatile solvents saturate the air cushion in your pipette by aspirating and dispensing the solvent repeatedly before aspirating the sample When the liquid temperature is different from t...

Page 11: ...nap cap and screw cap microtubes Description Part Number JIMMY Microtube Opener set of 3 F144983 For P5000 and P10mL models disposable filters maintain the pipette s original accuracy and precision by...

Page 12: ...P200 MyPIPETMAN FP070142 Metal Ejector Dark Blue P1000 MyPIPETMAN FP070151 Metal Ejector White P1000 MyPIPETMAN FP070152 PLASTIC TIP EJECTORS Model Part Number Plastic Ejector Dark Blue P2 P10 MYPIPET...

Page 13: ...Holder Dark Blue P1000 MyPIPETMAN FP070281 Tip Hholder White P1000 MyPIPETMAN FP070282 Tip Holder Dark Blue P5000 MyPIPETMAN FP070291 Tip Holder White P5000 MyPIPETMAN FP070292 Tip Holder Dark Blue P...

Page 14: ...place the tip holder S Connecting nut is loose Tighten connecting nut S M Damaged or corroded piston Return pipette to supplier S Improper repair or assembly Refer to MAINTENANCE S Pipetted volume is...

Page 15: ...pipette to the maximum volume given in the specifications and pre wet the tips 3 Aspirate the set volume from a reagent reservoir of distilled water 4 Maintain the pipette in the vertical position and...

Page 16: ...or wipe them with a soft cloth or lint free tissue soaked with the cleaning solution 3 Rinse the components with distilled water 4 Wipe the entire pipette with a soft cloth or lint free tissue soaked...

Page 17: ...e being immersed in decontamination solution 1 Disassemble the pipette refer to MAINTENANCE 2 Immerse tip ejector tip holder and connecting nut in the decontamination solution 3 Disassemble and degrea...

Page 18: ...p ejector along the tip holder 3 Clip the tip ejector on the body of the pipette Changing the Tip Holder for Single Channel 1 Push the ejection button 2 Turn the tip ejector clockwise 3 Slide and remo...

Page 19: ...p holder from the upper part refer to Figure 11 1 Remove the piston assembly O ring and the seals refer to Changing the Tip Ejector for Single Channel 2 Leave the piston exposed clean it with isopropa...

Page 20: ...MAN FP070105 Window 3digit Green MyPIPETMAN FP070106 Lubricant Tube for PIPETMAN 1g 5440011070 NOTICE For pipette service and maintenance or for extra spare parts please contact your local Gilson Serv...

Page 21: ...ilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipet...

Page 22: ...ilson Inc All rights reserved Specifications subject to change without notification errors excepted www gilson com contactus LT801622 D 2022 Gilson Inc All rights reserved Specifications subject to ch...

Reviews: