background image

California Bearing Ratio (CBR)

ASTM D1883; AASHTO T 193

HMA-684

6,000lbf load ring, dial indicator, piston

HMA-685

10,000lbf load ring, dial indicator, piston

HMA-685D 10,000lbf load cell, LVT, readout box, piston

Soil Cement

ASTM D1632, D1633

HMA-687

10,000lbf load ring, dial indicator, 4 in platen

HMA-687D 10,000lbf load cell , LVT, Readout box, 4 in platen

Triaxial

ASTM D2850, D4764; AASHTO T 296, T 297

HMA-686

1,000lbf load cell, dial indicator

HM-413

1,000lbf load cell, LVT, Readout box

Unconfined Compressive Strength

ASTM D2166; AASHTO T 208

HMA-681

500lbf load ring, dial gauge, plastic discs

HMA-683

1000lbf load ring, dial gauge, plastic discs

HMA-683D 1,000lbf load cell, dial gauge, plastic discs

COMPONENT SETS

3.  Connect to a properly grounded power supply with correct 

voltage and amperage to output.

4.  Adjust and level the crossarm to the appropriate height.

5.  Install required Component Set and any accessories required 

for testing.

CONTROLS

1.  Main Power On/Off: Includes an indicator light to show 

when the power is on

2.  Platen Direction: Three-position switch (up, down and off) 

includes built-in hesitation to prevent motor damage when 

reversing direction

3.  Limit Lights: Indicates when upper and lower limits are at-

tained

4.  Strain Rate: Three-segment thumbwheel selector (strain 

rate calibration chart included)

5.  Jog Button: Increases the speed of platen to the maximum 

speed of the load frame jack screw (0.1in/min) when the 

switch is pushed to the up position.  

OPERATING INSTRUCTIONS

  

WARNING!

Disconnect and lock out power supply when servicing the 

unit or installing components.

1.  Read all safety and operating instructions before operating 

the unit.

2.  Power the unit with the Main Power switch (indicator light 

will show that the unit is on).

3.  Based on the calibration chart (included), set the desired 

strain rate for the test using the three-segment thumbwheel 

selector.

4.  Use the Platen Direction and High-Speed Jog switches to ad-

just the direction and speed of the platen for testing.

5.  Center the Platen Direction switch to the off position when 

the test is complete.

6.  Refer to ASTM and/or AASHTO test methods for specific test 

instructions.

Accessories

HMA-94

Rolling Load Frame Cart

Reviews: