Gilson Maximum Perimissible Errors
Chapter 7
SPECIFICATIONS
PLATEMASTER is a high quality 96-channel manual pipetting system that offers
excellent accuracy and precision. The values given in the “Gilson maximum permissible
errors” table below were obtained using PIPETMAN DIAMOND Tips. These values are
guaranteed only when genuine PIPETMAN DIAMOND Tips are used.
PLATEMASTER P20 is certified with PIPETMAN DIAMOND Tips D200 and AmpliPur
Expert Tips.
PLATEMASTER P220 is certified with PIPETMAN DIAMOND Tips D300 and AmpliPur
Expert Tips.
Each PLATEMASTER is inspected and validated by qualified technicians in accordance
with the Gilson Quality System.
The adjustment is carried out under strictly defined and monitored conditions
(ISO 8655-6).
VOLUME*
GILSON
ISO 8655
SYSTEMATIC
ERROR
RANDOM
ERROR
SYSTEMATIC
ERROR
RANDOM
ERROR
PLATEMASTER P20 (0.5 — 20 µL)
(P/N F110761)
Min
1
±0.12
≤
0.1
±0.4
≤
0.20
10
±0.12
≤
0.1
±0.4
≤
0.20
Max
20
±0.2
≤
0.18
±0.4
≤
0.20
PLATEMASTER P220 (2 — 220 µL)
(P/N F110762)
Min
2
±0.12
≤
0.15
±8.0
≤
3.0
5
±0.25
≤
0.175
±8.0
≤
3.0
20
±0.4
≤
0.3
±8.0
≤
3.0
100
±1
≤
0.6
±8.0
≤
3.0
200
±1.6
≤
0.8
±8.0
≤
3.0
Max
220
±1.8
≤
0.8
±8.0
≤
3.0
*All Values in Microliters
S
PE
C
IF
IC
A
TIO
N
S
16
SPECIFICATIONS
|
PLATEMASTER®
Summary of Contents for F110761