IOM Sampling with the Gilian Gil Air Plus Pump Procedure
Calibrating GilAir Pump with IOM Sampler
To turn pump on, press the check mark/power button until screen lights up.
Check the flow rate:
1. Press the check mark/power button.
2. Use down arrow to scroll
to “Flow Set.”
3. Adjust the flow using the +/- buttons to set flow rate to
2000 cc/min (2 L/min)
.
After pump has charged:
1. Pull pump off charger.
2. Press check mark/power button.
Photo 1. Gilian Pump Calibration Setup
3. Use
“down” arrow until indicator is on “Calibrate.” Press check mark/power
button.
Allow the pump to run in calibrate mode for about five minutes.