Giga-tronics GT-1000B Microwave Power Amplifier
Operation
Operation Manual, Part Number 34672-001, Rev. A, May 2013
Page 11
Chapter 3
Operate the GT-1000B
3.1
Operating Safety and Instructions
This section describes how to operate the GT-1000B Microwave Power Amplifier. Equipment and personal
safety information is also provided.
DO NOT EXCEED AN INPUT LEVEL OF +17 dBm INTO THE GT-1000B.
EXCEEDING THIS LEVEL CAN DAMAGE THE GT-1000B MICROWAVE POWER
AMPLIFIER.
WHEN ENERGIZED, THE GT-1000B IS CAPABLE OF SUPPLYING POWER THAT
CAN CAUSE DAMAGE OR INJURY. TAKE THE FOLLOWING PRECAUTIONS TO
ENSURE SAFE SETUP AND OPERATION:
•
Verify that all cables, connectors, and equipment connected to the GT-
1000B are in good condition.
•
Do not make connections to equipment while the output of any item of
equipment is energized.
Table 6: Operate the GT-1000B
Operate the GT-1000B
Step
Action
1.
Verify that the output of the signal source driving the amplifier is NOT energized before
continuing to the next step.
2.
Connect the equipment to the GT-1000B according to your application.
NOTE:
Verify all mating connectors are 50 Ohm, and that they are in good condition.
3.
Apply AC power to the GT-1000B by turning the front panel power switch ON. Observe that the
POWER ON indicator on the front panel illuminates (see Figure 1 on page 4)
4.
Verify that the output level of the signal source is set below +17 dBm.
5.
Energize the output of the signal source.
6.
NOTE:
The GT-1000B has a Remote RF Off input. This input may be used to enable or disable the
RF output. Open circuit = RF On. Short circuit to ground = RF Off.
7.
NOTE:
Do not exceed the safe limits of the output power from the GT-1000B into the load. The
load should be rated for at least 30 Watts CW.
End of Procedure
CAUTION
WARNING
!