The PR command resets the 8650A to it’s default settings, leaving the user settings intact as
Previous
Settings
if they were different from the default settings. This command does not function while in the
SWIFT, FMOD or CM modes.
Alternatively, the IEEE 488.2 command *RST also resets the 8650A to it’s default settings, and
functions in the SWIFT, FMOD and CM modes. These modes must be re-entered either over the GPIB
or from the front panel.
The preset conditions of the instrument are outlined in Table 3-15.
Notes:
1.
The default measurement mode depends on the sensor type. For a CW sensor, the default is CW. For a modulation sensor,
the default is MAP. For a peak sensor, the default is INT TRIG. Regardless of the sensor type, all advanced features are
turned off.
2.
There is a slight difference between the preset conditions as set by a remote command, and as set from the front panel
menus. This difference has to do with measurement units. If the 8650A is preset from the front panel, this sets the
measurement units to dBm in all cases. If the 8650A is preset over the bus, this sets the measurement units to Watts in the
case of HP438 emulation, and has no effect at all in the case of HP436 emulation; otherwise, it sets the measurement units
to dBm.
This distinction is made to accommodate differences between emulations for remote programming purposes without
affecting the benchtop user.
Table 3-15: Preset (Default) Conditions
Sensors
(All parameters apply
to sensor A & sensor B)
General
Parameter
Condition
Parameter
Condition
Cal Factor
100.0%
Sensor Selection
Sensor A
Offset
0.00 dB
Calibrator
Off
Filter
AUTO
Default Sensor Prefix
Sensor A
Range
AUTO
Resolution
2 (0.01 dB)
Low Limit
0.000 dBm
Limits Checking
Off
High Limit
0.000 dBm
Max/Min
Off
Frequency
50 MHz
REL
Off
Duty Cycle
OFF, 1.000%
Trigger Mode
TR3
Measurement Mode
See Note 1
Group Execute Trigger Mode
GT2
Display Function
Display
Enable
Peaking Meter Mode
Status
Pulse Sensor Mode
Internal
Trigger
Measurement Units
See Note 2
Sound
On
Analog Output
Off
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...