Remote Operation
Manual 31470, Rev. E, April 2001
3-99
3.18.3
PAP Mode
Commands which specify the pulse average power measurement mode are based on the PAP function
code:
Syntax
:
PAP [A or B]
[A or B] specifies Sensor A or Sensor B.
These commands will work only with a modulated sensor. Possible GPIB entry errors: 60/61 (uncali-
brated or missing sensor A/B), 62/63 (not a modulated sensor, or two sensor operation active, A/B).
Examples
:
OUTPUT 713;PAP A
! select PAP mode for sensor A
OUTPUT 713;PAP B
! select PAP mode for sensor B
3.18.4
BAP Mode
Commands which specify the burst average power measurement mode are based on the BAP function
code:
Syntax
:
BAP [A or B]
[A or BE] specifies Sensor A or Sensor B.
These commands will work only with a modulated sensor. Possible GPIB entry errors: 60/61 (uncali-
brated or missing sensor A/B), 62/63 (not a modulated sensor, or two sensor operation active, A/B).
Examples
:
OUTPUT 713;BAP A
! select BAP mode for sensor A
OUTPUT 713;BAP B
! select BAP mode for sensorB
3.18.5
Peak Mode
The commands for Peak mode are discussed under separate headings for the 80350A sensors (see
Section
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...