Series 8650A Universal Power Meters
3-38
Manual 31470, Rev. E, April 2001
*STB?
Example
:
OUTPUT @Pwr_mtr;*STB?
! Query status byte register
Description
: This query returns status byte register result.
The SCPI status reporting structure includes the IEEE 488.2 Status Registers. Please
note that bit 5 of the OPERation Status Register is only used during BURSt and SWIFt
modes. NORMal mode does not use this function.
Figure 3-6: The SCPI Status Structure Registers
Not used
Not used
Not used
Not used
Not used
Waiting for BUS trigger BURSt/SWIFt
Not used
Not used
Not used
Channel 1 lower limit crossed
Channel 1 upper limit crossed
Channel 2 lower limit corssed
Channel 2 upper limit crossed
Not used
Not used
Not used
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
OPER
ation Status
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Not used
Not used
Not used
MAV
RQS
+
Operation Complete
Not used
Query Error
Device Dependent Error
Execution Error
Command Error
User Request
Power On
Status Byte
Standard Event
Status Register
Summary of Contents for 8650A Series
Page 4: ......
Page 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Page 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Page 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Page 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Page 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Page 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Page 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Page 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Page 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...