8003 Precision Scalar Analyzer
2-64
Manual 20791, Rev. C, June 2001
This softkey will bring up menu SY1-17. The available selections will be shown in standard
menu green. Non-available selections will be in dim, dark blue. In this menu, dim, dark blue
generally indicates that a sensor is not connected to an input.
To continue, select sensor A, B, or C. The display will change to show a calibration spreadsheet
for the selected sensor. The softkey menu will change to menu SY1-18.
[
PAGE UP]
First Level Menu SY1-18:
Accesses menu SY1-19. When [
PAGE UP
] is first pressed, the
spreadsheet will go to page 2 and the menu will change to SY1-19.
[
PAGE UP] and [PAGE DOWN]
Second Level Menu SY1-19:
Pages through the tables of calibration data to examine/change
entries. The bottom line of the screen displays new data as it is entered on the keyboard. To
change a data value in the table, move the yellow bar to the appropriate line using either the
spin knob or by repetitive entries on the UNITS keys (MHz/dBm or GHZ/dB). Enter the new
value using the keypad and the appropriate terminator key.
Subsequent operation of [
PAGE UP
] will continue to increment to the last page of data.
Use [
PAGE DOWN
] to return to an earlier page in the sequence.
[
FIRST PAGE] and [LAST PAGE]
Menu SY1-19:
These softkeys directly access the first and last pages, respectively, of the cali-
bration data.
[
ENABLE INPUT]
Menu SY1-19:
The default condition of the 8003 is for data input to be enabled. This can be
defeated using [
ENT/OFF
]. To re-enable, use [
ENABLE INPUT
].
[
SET TIME]
Menu SY1-18:
This softkey functions the same as [
SET TIME
] described on page 2-62.
[
CAL LOC]
Menu SY1-18:
Selects the location at which the calibration was performed. The choices are
shown in menu SY1-21. The CAL LOCATION is shown on page 1 of the sensor identification
spreadsheet.
[
PASSWORD]
Menu SY1-18:
Toggles the requirement for entry of a password before any changes can be
made to the sensor calibration stored in the EEPROM. The [
PASSWORD
] softkey accesses
menu SY1-22.
[
DEFINE PASSWORD]
Menu SY1-22:
Defines a password that users will have to enter to change the sensor calibra-
tion data stored in the EEPROM.
Summary of Contents for 8003
Page 3: ......
Page 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Page 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Page 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Page 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Page 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Page 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Page 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Page 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Page 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...