Giga-tronics 2500B Series Microwave Signal Generators
Special Modes
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 69
Table 27: Use Cable Correction
Use Cable Correction
Step
Action
Configure the 8650A/B series Universal Power Meter
1.
On the front of the 8650A/B series Power Meter, select the following:
1.
Meter Setup
2.
Config
2.
On the front of the 8650A/B series Power Meter, press the right arrow key to select RS-232; see Figure
Figure 34: 8650A/B Universal Power Meter; Front Panel Controls
3.
Select
Config.
4.
On the 8650A/B series Power Meter, set the RS-232 parameters as shown below. Use the Up/Down
Arrow keys to change the values of the parameters (see Figure 34).
Baud Rate; 38400
Data Bits; 8
Parity; None
Stop Bits; 1
5.
Press
OK twice to return to the power meter main menu.
Setup the Equipment
Softkey
Softkey
Softkey
Softkey