4.2
Open the adjustment menu
The sensor adjustment can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the display text
PR.OF
or
PR.SL
when switching on.
1.
Switch the device off.
2.
Hold the
down button
and press the
On/Off button
to switch on the device and
open the
Adjustment
menu.
3.
The display shows the first parameter. Release the
down button
.
Parameter
Values
Meaning
_
PR.OF
Zero point correction
0.0
No zero point correction
-5.0
…
5.0
Zero point correction [in selected display unit]
(e.g. ± 5.0 hPa)
PR.SL
Gradient correction
0.00
No gradient correction of the temperature
-5.00
…
5.00
Gradient correction in %
Formula:
Zero point correction:
Displayed value = measured value -
PR.OF
Gradient correction:
Display = (measured value -
PR.OF
) * (1 +
PR.SL
/ 100)