Software malfunction. 

8.

 

Reset (chap. 8.7) 

Individual motor does not work. 

1.

 

Motor-plug is not connected 
properly. 

2.

 

MCP is defective. 

3.

 

Control-box is defective. 

4.

 

Motor-cable or plug is defective. 

5.

 

Motor is defective. 

1.

 

Connect plug properly. 

2.

 

MCP must be replaced. 

3.

 

Control-box must be replaced. 

4.

 

Cord/plug must be replaced. 

5.

 

Motor must be replaced. 

 

15 

TECHNICAL DATA

 

15 

 

Characteristic 

Value 

Dimensions (length*width*height) 

205*85*50 cm 

Lying surface (length*width) 

205*85 cm 

Base (length*width) 

approx. 132*60 cm 

Height (no height adjustment) 

approx. 49 cm 

Inclination of headrest 

0

0

 to 40

0

 

Inclination of backrest 

0

0

 to 45

0

 

Inclination of legrest (thigh section) 

0

0

 to 23

0

 

Inclination of legrest 

0

0

-30

0

 

Lying surface lifting capacity 
(max. safe operational load) 

200 kg (centrally-positioned load on 

the centre of the lying surface; not 

permitted on the head and foot side) 

Backrest lifting capacity 
(max. safe operational load) 

75 kg 

Total weight 

approx. 90 kg 

Adjustment drive incl. controls 

Power of transformer (primary) 

230V AC/50Hz/200 VA max. 

Power of transformer (secondary) 

24 V DC 

Protection 

3 ampere AT 

Duty cycle  

10 % max. 2 min. uninterrupted 

operation/18 min. pause 

Power cord 

3 m, cord 

MCP 

Coiled cord 

Device protection class 

Overvoltage protection class 

Type B 

Protection  against  solid  and  liquid 
materials 

IP 54 

 

Ambient conditions for storing and transportation 

 

Temperature 

-10

C to +50

Atmospheric pressure 

700 to 1060 hPa 

Relative humidity 

20% to 90% at +30

0

 C non-condensing 

 

Summary of Contents for Evo Lounger

Page 1: ...L Wellness Couch Gharieni Evo Lounger Gharieni Group GmbH Gutenbergstra e 40 D 47443 Moers Phone 49 0 2841 88300 0 Fax 49 0 2841 88300 999 E mail service gharieni de Web www gharieni de Service Hotlin...

Page 2: ...peration 5 8 1 Pre set lying positions 5 8 2 Lock function 5 8 3 Headrest 6 8 4 Backrest 6 8 5 Inclination of the seat 6 8 6 Legrest 6 8 7 Reset 6 9 Lamp and tray 7 10 Cleaning 7 11 Maintenance 8 12 G...

Page 3: ...before you use the EVO LOUNGER for the first time Reading the complete IM is essential for its proper and confident use and will help you to operate the appliance to your entire satisfaction Our EVO...

Page 4: ...holstery is installed Open the top of the packaging carefully Do not use any sharp edged items which can damage the upholstery Do not lift the EVO LOUNGER but remove the packaging completely 5 BEFORE...

Page 5: ...ressing all moving operation will stop immediately 8 1 Pre set lying positions Set one of the electrically adjustable lying positions illustrated on the buttons by pressing the relevant button 8 2 Loc...

Page 6: ...the seat You can adjust the inclination of the seat section thigh section to knee bend section electrically Adjust the inclination of the seat section with the relevant button of the MCP 8 6 Legrest...

Page 7: ...g it if necessary Fig 3 10 CLEANING 10 Clean the housing and fittings with a damp cloth or sponge and a little bit of soap The standard upholstery surface of the EVO LOUNGER is made of PU polyurethane...

Page 8: ...ollowing conditions only 1 Repairs and modifications must be carried out by Gharieni Group GmbH or a person authorized by Gharieni Group GmbH 2 The electrical installation of the room the EVO LOUNGER...

Page 9: ...ystyrene pieces etc should be kept away from children CHOKING HAZARD The EVO LOUNGER is made of metal and plastic components electronic components electric cables Dispose of the separate components in...

Page 10: ...grest thigh section 00 to 230 Inclination of legrest 00 300 Lying surface lifting capacity max safe operational load 200 kg centrally positioned load on the centre of the lying surface not permitted o...

Page 11: ...H Manufacturer Gutenbergstra e 40 D 47443 Moers Germany declare that the product EVO LOUNGER Wellness couch Type or model name is produced in accordance with the requirements specified in Low Voltage...

Page 12: ...12 NOTES Gharieni Group GmbH Gutenbergstra e 40 D 47443 Moers...

Reviews: