Signet 9900 Conductivity Transmitter 

Standards and Approvals

Module Installation

Optional:
Relay Module
(Panel Mount 
Only)

Conductivity/

Resistivity

Module

Optional:
H COMM Module
or Modbus Module

Base Unit

H COM

DC Power

Loop V

oltage

Rating:

  5A
 V

A

C    

~

5A

 30 VDC 

׽׽

3-9900.394

Cond/Res Module

3-9900.393  Relay Module

 

Minimize handling of the plug-in modules to reduce the 
possibility of damage due to ESD.

 

Handle modules by the edges. 
Never touch any exposed circuitry or contacts.

 

Wear an anti-static wristband or stand on an anti-static mat, 
or keep one hand touching a properly grounded pipe or other 
piece of properly grounded metal when handling modules.

CAUTION 

Avoid Electrostatic Discharge (ESD).

CAUTION

Exercise care when installing modules.

Do not bend connecting pins.

CAUTION

LOOP as well as DC power MUST be removed 

BEFORE installing or removing ANY Module.

If the 9900 Base Unit will be mounted in a panel, the plug-
inmodules may be installed either before or after the base unit 
is mounted. If the 9900 Base Unit will be mounted using the 
wall mount accessory kit (3-9900.392), install plug-in modules 

fi

 rst. If installing both the Conductivity/Resistivity (Cond/Res) 

and the H COMM Modules, install the H COMM Module 

fi

 rst, 

then the Cond/Res Module on top of it.

To install modules:

Remove power from the 9900. Carefully align pins and
connectors (do not bend connecting pins) and push module 

fi

 rmly into place, then attach with screw(s) (except H COMM 

Module).

Standards and Approvals

• 

CE, UL, CUL, WEEE

• RoHS 

Compliant

• Lloyd's 

Register

• 

Manufactured under ISO 9001 for Quality, ISO 14001 for 

Environmental Management and 

OHSAS 18001 for Occupational Health and Safety.

    China RoHS (Go to gfsignet.com for details)

 Declaration of Conformity according to FCC Part 15 

• 

This device complies with Part 15 of the FCC rules. 

Operation is subject to the following two conditions: 

(1) This device may not cause harmful interference, and 

(2) This device must accept any interference received, 

including interference that may cause undesired operation.

Summary of Contents for Signet 9900

Page 1: ...085 0 01 cm 1 316 SS electrode Sanitary Tri clamp fl ange 1 to 1 in 3 2820 1 198 844 000 0 1 cm 1 316 SS electrode in threads in line only 3 2821 1 198 844 001 1 0 cm 1 316 SS electrode in threads in...

Page 2: ...l warranty and non warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales office or distributor Product returned without a Servic...

Page 3: ...the plug inmodules may be installed either before or after the base unit is mounted If the 9900 Base Unit will be mounted using the wall mount accessory kit 3 9900 392 install plug in modules first I...

Page 4: ...s provided to help guide the cutting process De burr and smooth the opening with a file 1 The panel mount transmitter is designed for installation using a DIN punch Recommended clearance on all sides...

Page 5: ...and mechanical damage Seal the cable entry points to prevent moisture damage Only one wire should be inserted into a terminal Splice double wires outside the terminal or use appropriate wire ferrule...

Page 6: ...o be used for system power 10 8 to 35 2 VDC NOTE Backlight LEDs Modbus and optional Relay Module do not operate on loop power Any connected sensors or sensor electronics that cannot operate on loop po...

Page 7: ...s Module which Provides filtering and signal conditioning The standard sensor cable length can be extended up to 30 m 100 ft Loop Voltage DC Power Cond Res Module SHIELD TEMP WHT ISO GND BLK SIGNAL RE...

Page 8: ...PNP Style Wiring The 9900 Open Collector R1 output provides high speed switching capability Signal frequencies can reach 400 pulses per minute The 9900 Open Collector R1 output connection is dependen...

Page 9: ...d is often required to start editing Once entered correctly this password will not be needed for subsequent edits However once the menu system is exited the password will again be required when edit m...

Page 10: ...hanical Relay R2 Indicator LED Backlight Sensor do not block Warning LED Bar Graph Units of Measure GPM pH sec etc Value Label Menu Indication Menu Navigation Keys Mechanical Relay R3 Indicator LED Al...

Page 11: ...new value selection 6 Press or to select another menu item Repeat steps 3 5 as required 7 Press again to return to normal operation The menu is constructed in a loop so you can move forward and backw...

Page 12: ...display LOOKING FOR If no sensor is attached to the 9900 the words TYPE and COND RES are displayed If the 9900 does not identify your sensor type correctly use the and keys to select your sensor type...

Reviews: