ULTIMA series
ALARMS
NU_ULTIMA_02_EN_1612 51
3.
WARNINGS
After the Start icon is pressed, the system checks for any
problems and will notify the operator via warning messages.
Pressing the
icon will clear the warning.
HOT CHAMBER
The following message is displayed if the temperature inside the washing
chamber is above the safety temperature set point for door opening.
At the end of the cycle, let the washer chamber and contents cool
before opening the chamber door.
ACTIVE ALARME
At least one alarm is active
RECIPE ERROR
The selected recipe is not configured
ENTER A BATCH ID
Required batch ID is missing.
DOOR OPEN
Door is opened or unlocked.
ADDITIVE 1 LOW LEVEL
* optional for 810LX and 815 LX
The following message is displayed at the start of the cycle.
It is indispensable to replace the tank.
Consider checking the level of other product tanks in order to avoid
another alarm.
The warning for low level prohibits the equipment to start if not the
corrective action has been made to change chemical canister.
ADDITIVE 2 LOW LEVEL
* optional for 810LX and 815 LX
ADDITIVE 3 LOW LEVEL
** optional extra alkaline dosing pump
ADDITIVE 4 LOW LEVEL
** optional extra alkaline dosing pump
ADDITIVE 5 LOW LEVEL
** optional extra alkaline dosing pump
END OF CYCLE : OPEN THE DOOR
Successful cycle completed. System waiting.
Summary of Contents for 1300 LX
Page 1: ...ULTIMA series USER MANUAL...
Page 8: ......
Page 34: ......
Page 37: ...ULTIMA series ACCESS TO THE USER S MENU NU_ULTIMA_02_EN_1612 35...
Page 46: ...ULTIMA series ACCESS TO THE USER S MENU NU_ULTIMA_02_EN_1612 44...
Page 50: ......
Page 58: ......
Page 68: ......
Page 81: ......
Page 82: ......