Specifications
48
PhysioMem
®
PM 100 2G
77811011
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The
1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used be-
cause while it does not represent actual modulation, it would be worst case.
WARNING
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
WARNING
Use of accessories, transducers and cables other than those speci-
fied or provided by the manufacturer of this equipment, could result
in increased electromagnetic emissions or decreased electromag-
netic immunity of this equipment and result in improper operation.
WARNING
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm to any part of the device, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this
equipment could result.