25
LRR 1-50, LRR 1-51 - USA - Installation & Operating Manual -
850702-01
Operating the conductivity controller
Meaning of codes on the 7-segment display
Code Meaning
Indicated when rotary knob is turned clockwise:
AL.Hi Alarm High
MAX switchpoint, adjustable from 0.5 to 4999.5 ppm (1 to 9999 µS/cm)
SP
Set point
Adjustable from 0.5 to 4999.5 ppm (1 to 9999 µS/cm)
HySt Hysteresis
Reset hysteresis, adjustable from 1 to 25% of set point
FiLt
Filter
Switching the filter on/off (damping)
PW
Password
on = password protection is enabled oFF = password protection is disabled
Factory default setting 1902 (cannot be changed)
Indicated in parameterization mode
quit
Quit
Entry is not confirmed
done
Done
Entry is confirmed
Indicated in the event of errors
E.001
Error
Faulty temperature sensor, temperature reading too low (LRR 1-50 only)
E.002
Error
Faulty temperature sensor, temperature reading too high (LRR 1-50 only)
E.005
Error
Faulty measured value acquisition, reading too low
E.006
Error
Faulty measured value acquisition, reading too high
LRR 1-50 only
CAL Electrode
calibration
Electrode calibration, last reading
is shown
CF
Correction
factor
Adjustable from 0.05
to 5.000 in increments
of 0.001
inP Input for
Pt100
Temperature compensation
YES (no)
tC
Temperature
coefficient
tC 0.0 – 3.0% per °C,
adjustable in increments of 0.1
LRR 1-51 only
Sin.L
Lower end of measuring range, adjustable
from 0.0 - 0.25 - 25 - 50 ppm
(0 - 0.5 - 50 - 100 µS/cm)
Sin.H
Upper end of measuring range, adjustable
from 10 - 50 - 100 - 250 - 500 - 1000 -
1500 - 2500 - 3000 - 3500 - 5000 - 6000
ppm (20 - 100 - 200 - 500 -1000 -
2000 - 3000 - 5000 - 6000 - 7000 -
10000 - 12000 μS/cm)
Sout
Standardization of current output, adjustable between 5 and 5000 ppm
(10 and 9999 µS/cm)
Si
Flushing interval, adjustable from 0 to 24 hours in increments of 1h
Sd
Flushing time, adjustable from 1 to 4 minutes in increments of 1min.
tESt
Test
Test of output relays
Fig. 6
Rotary knob with
integrated
push-button
Red MAX LED
7-segment display
LED 1 for
valve opening
LED 2 for
valve closing