4 Assembly and setting into operation
Use of the set of tanks KAN 20 (optional):
- Connect the diode plugs of the level detectors to the distribution box and con
-
nect to the socket level detector (see fig. 3.4, pos. 4).
- Blank off the remaining diode sockets of the distribution box with the dummy
diode plugs enclosed.
Usage of other set of tanks without level sensor control
- Put the diode dummy plug on the connection of level sensor control (see ill.
3.4., pos.4).
4.3. Mains connection
Please check the nominal voltage on the type plate before connecting to the mains.
Should the nominal voltage deviate more than +/- 10%, then you have to contact
your service engineer.
- Make sure that the mains switch at the front of the instrument is in position "0"
- Connect the mains cable to the rear appliance plug first
- Connect the mains cable to the shockproof socket
4.4 Setting into operation
Attention when working with acids and alkali! Make sure you observe the safety
instructions concerning work with hazardous materials!
Always close the plexiglass doors of the distillation unit before operating!
1. Fill the set of tanks with chemicals:
- H2O distilled or demineralized
- NaOH 32 %
2. Turn on the water tap.
3. Before you set the Vapodest into operation make sure that the excess current
switch at the rear is in position 1.
4. Turn on the Vapodest 10 using the mains swich of the instrument.
5. The Vapodest 10 is ready to use