GERATHERM UniqueResc+ Instruction Manual Download Page 7

w

w

w

.g

er

at

h

er

m

.c

o

m

7

•  Do not operate the control unit in areas subject to 

explosion hazard!

•  Do not use Geratherm UniqueResc

+

 warming blan-

kets together with other heat sources (e.g. radiation 
lamps, incubators or hot water bottles)! This could 
be detrimental to the patient’s health.

•  Do not place the control unit directly on top of ECG 

or EEG equipment. Otherwise, faults may arise, e.g. 
the devices may produce false readings!

 
•  Before each use, check the surface of the war-

ming blankets for any visible damage! Disconti-
nue use of any damaged blankets immediately!

•  Put the Geratherm UniqueResc

+

 warming blankets 

completely on the patient. To prevent overheating, 
do not under any circumstances fold the equip-
ment!

•  Do  not  place  any  objects  on  top  of  the  warming 

blankets to avoid any localised overheating or heat 
accumulation.

•  It must be checked whether any fixture is required 

to keep the patient in place. If so, the same must be 
positioned underneath the warming blankets.

•  The temperature setting (only by model type R130) 

is the responsibility of the physician.

 
•  Do  not  use  Geratherm  UniqueResc

+

 on patients 

with impaired circulation! The manufacturer is 
not aware of any other contra-indications.

 
•  Whenever heat treatment is applied, the core body 

temperature of the patient must be monitored at re-
gular intervals. The length of the intervals shall be 
determined by the medical staff.

 
•  Geratherm  UniqueResc

+

 shall only be used under 

the supervision of the medical staff.

  Do not apply adhesive tape to the warming blan-

kets. This will lead to permanent soiling of the 
surface.

Summary of Contents for UniqueResc+

Page 1: ...www geratherm com Instruction manual PatientWarmingSystem M e d i c a l W a r m i n g S y s t e m s UniqueResc Warmth that saves lives...

Page 2: ...warming blankets by application of modern insulating material Noiseless no distribution of germs through air distur bance Simple to use Plug in forthwith warming Highly durable and easy to clean as t...

Page 3: ...ing up 14 5 4 Switching off the warming blankets 15 6 Shutting down 16 6 1 Shutting down after use 16 6 2 Shutting down in order to storage for a certain length of time 16 7 Messages and faults 17 pag...

Page 4: ...eets the requirements of Council Directive 93 42 EEC of 14 06 93 on Medical Devices and of the standard DIN EN 60601 1 Medi cal electrical equipment Part 1 General requirements for safety DIN EN 60601...

Page 5: ...patient from above The use of several blankets per patient is possible The flexibility of the blankets makes it possible to warm the patient even during complicated rescue operating terms Geratherm Un...

Page 6: ...rmally conductive materials as water snow and similar substances the patient s body temperature could cool down as long as the device is disconnected All of the Geratherm UniqueResc must not be operat...

Page 7: ...ces fold the equip ment Do not place any objects on top of the warming blankets to avoid any localised overheating or heat accumulation It must be checked whether any fixture is required to keep the p...

Page 8: ...www geratherm com 8 Complete warming blanket example of application 4 Description 4 1 Overview of the components and manual control elements 4 1 1 Type R130...

Page 9: ...splay lights up during operating 4 LED display ERROR lights up if a fault occurs 5 Display of the selected set temperature 6 SELECT button for switching over the set temperature 7 Safety class 8 Do no...

Page 10: ...www geratherm com 10 4 1 2 Type R150 example of application Complete warming blanket...

Page 11: ...splay flashes during heating up 3 LED display POWER display lights up duringoperating 4 LED display ERROR lights up if a fault occurs 5 Safety class 6 Do not dispose of in household waste 7 BF type ap...

Page 12: ...em possesses monitoring installation of hard ware as well as software A short tone after the unit is switched on indicates that the alarm system and the independent safety device have been automatical...

Page 13: ...y the patient on top of the Geratherm UniqueResc warming blankets Put the warming blankets on the patient in such a way that the control unit s window side is facing away the patient The warming blank...

Page 14: ...f the 3 LED s light up briefly All required heating system equipment checks are performed automati cally in this period of time At the same time a short tone signal announces the test of the acoustic...

Page 15: ...ing blanket from 20 C to 37 C is approx 5 minutes When executing R130 and the power supply is inter rupted the temperature is re set independently from the eventually preselected temperature to 41 C I...

Page 16: ...the warming blankets 6 2 Long termed shutting down of the Geratherm UniqueResc Prior to storing UniqueResc no maintenance is re quired The system may only be deposited dryly If ring connectors are oft...

Page 17: ...e service technician Due to high sensitivity of the sensors the ERROR LED could probably flash briefly ca 2 3 seconds during the heating process This is no system error and it is reset autonomously If...

Page 18: ...h the materials must first be tested on a test area Upon re quest Geratherm will be pleased to provide samples of the materials for such tests A suitable product is for example Incidur Spray manufactu...

Page 19: ...connecting cables re gularly for any damage Any damage to the surface coating due to long term intensive use or strong mechanical forces necessitates replacement by the manufacturer Damaged blankets...

Page 20: ...g safety related checks by the manufacturer or an authorised person every 24 months Checking the calibration of the systems is only required after the replacement of assembly modules and as part of th...

Page 21: ...ing blankets The warranty does not include the remedying of damage arising as a result of accidents disasters improper use or other external influences use of parts that do not belong to the product m...

Page 22: ...C Humidity to 95 Transport and storage Temperature 30 C to 70 C Humidity to 95 12 2 Type R150 Art No 8001110616 Nominal voltage 12V to 28 V DC Power max 100 VA Fuse 16A M delayed action 6x25 mm in th...

Page 23: ...nment Temperature 10 C to 40 C Humidity to 95 Transport and storage Temperature 30 C to 70 C Humidity to 95 12 4 Scope of delivery UniqueResc R130 Art Nr 8001110615 Warming blanket 50cm x 86cm with in...

Page 24: ...rmity of the medical product with Directive 93 42 EEC Theidentificationnumber0118standsfortheThuringian State Office for Measurement and Calibration Pay attention to operating instructions BF type dev...

Page 25: ...ial precautions as regards elec romagne tic compatibility EMC and must be installed and operated in compliance with the EMC information included in the accompanying documentation 2 A statement that po...

Page 26: ...B The device or system is suitable for use in all facilities including households and facilities that are directly connected to the public low voltage electri city supply network that supplies buildi...

Page 27: ...Differential mode 2 kV Common mode The quality of the line current should cor respond to that of a typical commercial and or hospital environment Voltage reductions short interruptions and voltage fl...

Page 28: ...ecommended distance Conducted HFIEC 61000 4 6 3 V effective value 150 kHz to 80 MHz 3 V effective value d 3 5 3 P 1 17 P Radiated HF IEC 61000 4 3 3V m 80 MHz to 2 5 GHz 10 V m d 3 5 10 P 0 35 P 80 MH...

Page 29: ...ate the electromagnetic environment around fixed HF transmitters an electromagnetic location check should be considered If the field strength measured at the location where the device or system is use...

Page 30: ...att Distance m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 17 P d 0 35 P d 0 70 P 0 01 0 12 0 04 0 07 0 1 0 37 0 11 0 22 1 1 17 0 35 0 70 10 3 70 1 11 2 21 100 11 70 3 50 7 00 For trans...

Page 31: ...www geratherm com 31 Manufacturer The manufacturer reserves the right to make technical modifications in keeping with technical advances...

Page 32: ...32 www geratherm com Geratherm Medical AG Fahrenheitstrasse 1 98716 Geschwenda Germany Phone 49 0 36 205 98 0 Fax 49 0 36 205 98 116 E Mail info geratherm com www geratherm com...

Reviews: