GERATHERM Medical AG wristwatch Instruction Download Page 2

 

When the measurement was finished, the 
display will show the measured pressure 
and pulse value back and forth. Then, 
press the power key to turn off the unit. 
Otherwise the unit will turn off 
automatically in 24 seconds later. 
 
 
 
 
 
 
 
 
2. Recall the memories 
 
Press and release the MEMORY key to 
recall the no. 1 memory. The display will 
show the memorized pressure and pulse 
value back and forth. 
 
 
 
 
 
 
 
 
 
 
 
 
 
Press and release the MEMORY key 
again to recall the no.2 memory. Press 
and release repeatedly to read the other 
memorized pressures.  
 
3. Clock setting 
 
To enter the setting mode 
by pressing and holding the 
memory key lasting for over 
3 seconds. 
The display will show a 
blinking year  firstly. 
Press the power key to adjust 
the year when it is blinking. 
Secondly, to press and 
release the memory key, 
the “month” will appear and 
blink to be adjusted, and then 
“date”, “hour” and “minute” will 
follow up as you press and 
release the memory key again. 
 
 


To adjust the clock 

 

by pressing the memory key to adjust the 
values of the “month”, “date”, “hour” 
and “minute” when they are blinking 
 

Care and Maintenance 

 
 
Do not drop the 
 unit. 
 
 
 
Do not modify or  
disassemble the 
body or the wrist 
cuff. 
 
 
Do not bend or 
twist the wrist 
cuff. 
 
 
 
Use a moistened  
cloth with water or 
neutral detergent to 
clean the body and 
then wipe it dry. 
 
 
Avoid thinner  
and benzine. 
 
 
 
 
Keep the unit in 
an suitable place 
avoiding high 
temperature, 
direct sunlight, 
high moisture, 
and much dust. 
 
 
Remove the  
batteries, if the 
unit will be not 
used for a long 
time. 
 
Do not press the 
POWER key when 
 the cuff has not  
been well wrapped. 

 

Error Message 

 
Error Marks 

Correction 

 
 

 

The pressure measured 

                               was lower than 20mmHg. 
 

 

Please measure again. 

 
 
 

 

The pressure measured 

                                Was higher than  
                               300mmHg. 
 

 

Please measure again. 

 

The pumping pressure is 
higher than 300 mm HG. 
Please measure again. 

 
 
 

 

Please replace the 

 

 

2 batteries. 

 
 
 

 

Check or Replace 

 

 

the 2 batteries and  

 

 

try again. 

 
 
Specification 
 
Name and model  Wrist Blood Pressure 
 

Monitor KP-6120,KP-6130 

Display system 

Liquid Crystal Display 

Measuring method Fully fuzzy logic 
Power Source 

2 Alkaline “AAA” type batteries 
(1,5V) 

Measuring range 

20-300 mmHg 

 

(Cuff Pressure) 

 

40-200 Pulse/minute 

 

(Pulse Rate) 

Accuracy 

3mmHg  

 

 (Cuff Pressure) 

 

5% of Readings 

 

(Pulse Rate) 

Inflation 

Automatic inflation 

 

(Air Pump) 

Deflation 

Deflation valve 

Rapid air release 

Electric release valve 

Memory 

7 meAverage value 

Clock 

Year, Month, Date, Time 

Error Message 

“Err LL” means the pressure 
was lower than 20 mmHg. 
“ErrUU” means the pressure 
was higher than 300 mmHg. 

Low battery 

Replace the batteries if the 
indicator appears. 

Power OFF 

Automatically after 150 
seconds of non use. 

 

Battery Life 

Approximately 250 
measurements. 

Operating 

+ 10°C to + 40°C 

environment 

less than 85% RH 

Dimensions 

78(W) X 74(D) 

                               X 30 mm (H) 
Weight 

Approx. 130g 

 

(including 2 batteries) 

 
 
 
 
 
 
 
 
Note : Specification are subject to  
change without notice for 
improvement. 
 
The blood pressure monitor devices complies 
With the European regulations (EU council 
directive 93/4

2/EEC) and bears the CE mark “CE 

0118”. The quality conforms to the following 
standards EN 60601-1-2 (Medical electrical 
equipment - Part 1: General requirements for 
safety); EN 1060-1 EN 1060-3 ( Non-invasive 
sphygmomanometers - Part 1: General 
Requirements. Part 3: Supplementary 
Requirements to electro-mechanical blood 
pressure measuring systems). 
Production according to EN ISO 13485. 
If used in a professional context, the thermometer 
should be inspected every two years by an 
authorised laboratory. 
 
 
Produced for  
 

Geratherm Medical AG 
Fahrenheitstraße 1 
98716 Geschwenda 
Germany 
 

 
 

 

 

No display 