User Manual
Ergostik
ERGOSTIK_MAN_ENG_V1.2.2_REV01.DOCX
Page 24
4.
Technical Data Ergostik
Modifications are not permitted with this device
Technical Data:
Dimensions:
Weight:
210 mm x 175mm x 75mm (L x w x h)
1120 g
Electrical data:
IP protection type:
Protection class:
Classification according to
MDD:
Application component type:
PC interface:
Power supply:
Power consumption:
IPX0 IEC 529
II
IIa 93/42/EWG date June 14th 1993 14.06.93
Appendix IX
BF according to VDE 0750 (DIN EN 60601-1)
USB 2.0
12V max. 5A
<60 VA
Flow:
Flow sensor:
Measuring principle:
Measuring range:
Measuring range ventilation:
Resistance:
Effective dead space:
Flow resolution:
Sample rate:
Accuracy:
Ergoflow
Differential pressure
± 16l/s
0 – 300l/min
<0.12kPa/(l/s) <15l/s
<20ml
< 1ml/s
125 Hz
±3% or 20ml/s
Volume:
Measurement range:
Accuracy:
0 – 20L
± 3% or 50mL
O2 Analyser:
Measurement principle:
Measurement range:
Accuracy:
Resolution:
T
90
:
Electro Chemical cell
1 – 100% O2
0.1%
0.1%
< 100 ms
CO2 Analyser:
Measurement principle:
Measurement range:
Accuracy:
Resolution:
T
90
:
Infra-red absorption
0 – 13% CO2
0.1%
0.1%
28 ms
Operating conditions:
Temperature:
Air humidity
+15°C to +40°C
10 to 90% (non-condensing)
Storage and Transport
conditions:
Temperature:
Air Humidity:
Explosive conditions:
-10°C to +60°C
0 to 95% (non-condensing)
Device should not be used in explosive or
flammable atmospheres
Minimal requirements PC
system:
Norm:
Processor:
RAM Memory:
Hard disk:
Monitor:
Interface:
Operating system:
DIN EN 60950
Pentium III compatible or higher
1Ghz or higher recommended
500 MHz minimum
512 MB RAM or higher
192 MB RAM minimum
1.5 GB or greater recommended
600 MB minimum free space required
XGA (1024 x 768) or higher
USB 2.0 recomended
USB 1.1 minimum
Windows XP SP2 or higher