Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Version: 6 | Release date: 02 August 2022
Page 71
Possible danger to life.
Reason: Electric shock. Misdiagnosis caused by
measurement error. For this: Ensure that the used
components are undamaged and that you are working
in a careful way! Therefore:
•
Prior to each use, visually inspect the total
system (housing, connectors, etc.) for any
damages!
•
If there are any damages, do not operate the
system. The damaged parts must be replaced
or repaired properly!
Reason: Electric shock. Triggering malfunctions of
Ergostik Blueflow. Unqualified user / operator can
detect sources of failures too late or even cause them.
Therefore:
•
All maintenance and servicing work may only be
performed by specialised personnel who have
been authorised by the manufacturer!
8.2.1
Checking / Exchanging of Tubes and Cables
All parts of Ergostik Blueflow should be checked for visible
mechanical damage (cracks, tears) each day. If damage is
ascertained, the corresponding component must be replaced.
For replacing components see:
•
Chap. 4.1.3 “System Construction and Electrical
Safety“
•
Chap. 6.2 “Recommissioning after Servicing / Cleaning
Work“