Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Page 48
Version: 6 | Release date: 02 August 2022
4.1.2.3
Pressure Reducer
For safe use, a suitable pressure reducer for the calibration gas
cylinder is required. Please note the following:
Attention
Ergostik Blueflow could be damaged
Reason: Working pressure too high. Therefore:
•
Only use a pressure reducer that meets the
specifications defined by the manufacturer and
provides an outlet pressure of maximum 5 bar!
The manufacturer has approved a suitable pressure reducer as
an optional extension. You can get it from your authorised
specialist retail partner.
•
See also chap. 2.2.1 "Original Spare Parts /
Accessories / Optional Extensions".
4.1.3
System Construction and Electrical Safety
The following instructions are intended for safe handling of the
entire system, taking into account the electrical safety concept of
the Ergostik Blueflow.
The system may be set up only by an authorised specialist retail
partner.
It is essential to note: Anyone who combines additional
devices or medical devices or unauthorised or non-original
components / spare parts / consumables with existing
medical, electrical equipment or systems, and this
combination is used by third parties or this combination is
placed on the market, will legally become a producer of a
system or a procedure pack. In any case the assembler of a
system is therefore responsible for compliance with the
requirements placed on the system by the relevant,
harmonised standards and the additional national and
international standards and guidelines in the currently valid
versions!